The accuracy and outcome of capillary blood glucose (CBG) monitoring as routinely performed by nursing staff were assessed. The sample consisted of 160 readings conducted by 93 nursing staff members in four hospitals; 19% of the readings deviated from simultaneous laboratory results by >20%, and deviations resulted in altered responses to standing orders in 26 patients (17%). There was no statistically significant difference between the total variation attributed within and between nurses, possibly indicating that all nurses should be given the same intensity of follow-up training rather than targeting those who perform incorrectly on 298 DIABETES CARE, VOL. 12, NO. 4, APRIL 1989 SHORT REPORTS sample tests. Nurses in the one hospital that required certification before CBG monitoring had significantly less deviation from the laboratory standard than the other three hospitals. Although these data do not provide direct evidence that the certification program increased accuracy, this seems a logical conclusion. More study is needed to determine the most costeffective type of training and follow-up.
Skip Nav Destination
Article navigation
Short Reports|
April 01 1989
Accuracy of Nurses in Performing Capillary Blood Glucose Monitoring
Patricia A Lawrence, RN, MA;
Patricia A Lawrence, RN, MA
School of Nursing, University of North Carolina
Chapel Hill, North Carolina
Search for other works by this author on:
Mary C Dowe, RN, EdD;
Mary C Dowe, RN, EdD
School of Nursing, University of North Carolina
Chapel Hill, North Carolina
Search for other works by this author on:
Elizabeth K Perry, RN, BSN;
Elizabeth K Perry, RN, BSN
School of Nursing, University of North Carolina
Chapel Hill, North Carolina
Search for other works by this author on:
Susan Strong, RN, MSN;
Susan Strong, RN, MSN
School of Nursing, University of North Carolina
Chapel Hill, North Carolina
Search for other works by this author on:
Gregory P Samsa, PhD
Gregory P Samsa, PhD
School of Nursing, University of North Carolina
Chapel Hill, North Carolina
Search for other works by this author on:
Address correspondence and reprint requests to Patricia A. Lawrence, RN, MA, School of Nursing, CB 7460, Carrington Hall, University of North Carolina, lf-moni- Chapel Hill, NC 27599.
Citation
Patricia A Lawrence, Mary C Dowe, Elizabeth K Perry, Susan Strong, Gregory P Samsa; Accuracy of Nurses in Performing Capillary Blood Glucose Monitoring. Diabetes Care 1 April 1989; 12 (4): 298–301. https://doi.org/10.2337/diacare.12.4.298