A novel enzyme-linked immunosorbent assay (ELISA [Dialbumin]) for rapid office measurement of microalbuminuria was evaluated and its performance compared with that of a commercially available radioimmunoassay (double-antibody albumin). Urine samples containing between 0.75 and 1800 μg/ml of albumin were obtained from 31 diabetic patients and assayed by both methods. A comparison of the paired values obtained from the two methods gave a correlation coefficient of >0.99. The Dialbumin assay, which used detachable eight-well strips (1 strip/sample), 10-min incubation, tap water wash, and a 2-min color development step, was read on both an ELISA reader and a hand-held analytical device (Acc-U-Dial) designed specifically for this test. The findings of this study indicate that the Dialbumin assay, usedin conjunction with the Acc-U-Dial device, affords a rapid, convenient, and sensitive method for quantitative determination of a broad range (0.3–1280 μg/ml) of urinary albumin levels in the office setting.
Skip Nav Destination
Article navigation
Technical Article|
October 01 1990
Evaluation of New Rapid Office Test for Microalbuminuria and Its Comparison to Fully Quantitative Radioimmunoassay
R Glenn Neuman, MS;
R Glenn Neuman, MS
Exocell and the Department of Medicine, University of Pennsylvania
Philadelphia, Pennsylvania
; and the Department of Endocrinology, University of Parma
Parma, Italy
Search for other works by this author on:
Luigi V Bonomini, MD;
Luigi V Bonomini, MD
Exocell and the Department of Medicine, University of Pennsylvania
Philadelphia, Pennsylvania
; and the Department of Endocrinology, University of Parma
Parma, Italy
Search for other works by this author on:
Seth N Braunstein, MD
Seth N Braunstein, MD
Exocell and the Department of Medicine, University of Pennsylvania
Philadelphia, Pennsylvania
; and the Department of Endocrinology, University of Parma
Parma, Italy
Search for other works by this author on:
Address correspondence and reprint requests to R. Glenn Neuman, MS, Exocell, Inc., 3508 Market Street, Room 420, Philadelphia, PA 19104.
Diabetes Care 1990;13(10):1069–1073
Article history
Received:
September 08 1989
Revision Received:
May 09 1990
Accepted:
May 09 1990
PubMed:
2209304
Citation
R Glenn Neuman, Luigi V Bonomini, Seth N Braunstein; Evaluation of New Rapid Office Test for Microalbuminuria and Its Comparison to Fully Quantitative Radioimmunoassay. Diabetes Care 1 October 1990; 13 (10): 1069–1073. https://doi.org/10.2337/diacare.13.10.1069