This paper describes a miniature precision peristaltic pump and a silicone rubber reservoir developed for the infusion of insulin in diabetic subjects. The ability of the 100-ml reservoir to maintain sterility of its contents over 26 refilling cycles (48 days) is demonstrated. Accelerated bench tests on 10 pumps over 30 days indicate an accuracy of ± 1% on flow rate whether measured over 10 min or 24 h and with or without 1 atm of outlet back pressure. Comparison of calibrations made before and after 58 animal and clinical tests of 6–71 days′ duration showed a percentage difference in calibration that varied with a mean ± SD of 0.05 ± 3.09%. Microscopic examination of pump tube samples after 21 days of clinical use indicated redistribution of surface material without evidence of particle spallation. Samples subjected to accelerated wear equivalent to 620 days of clinical use showed extensive luminal surface modification and particulate deposits on downstream filters. These were identified by X-ray spectroscopy as silicon compounds.
Skip Nav Destination
Article navigation
Proceedings of the Kroc Foundation International Conference|
March 01 1980
For Insulin Infusion: A Miniature Precision Peristaltic Pump and Silicone Rubber Reservoir
W S Jackman;
W S Jackman
Hospital for Sick Children and the Toronto General Hospital
Toronto, Canada
Search for other works by this author on:
W Lougheed;
W Lougheed
Hospital for Sick Children and the Toronto General Hospital
Toronto, Canada
Search for other works by this author on:
E B Marliss;
E B Marliss
Hospital for Sick Children and the Toronto General Hospital
Toronto, Canada
Search for other works by this author on:
B Zinman;
B Zinman
Hospital for Sick Children and the Toronto General Hospital
Toronto, Canada
Search for other works by this author on:
A M Albisser
A M Albisser
Hospital for Sick Children and the Toronto General Hospital
Toronto, Canada
Search for other works by this author on:
Address reprint requests to A. M. Albisser, The Hospital for Sick Children, 555 University Avenue, Toronto M5G 1X8, Canada.
Citation
W S Jackman, W Lougheed, E B Marliss, B Zinman, A M Albisser; For Insulin Infusion: A Miniature Precision Peristaltic Pump and Silicone Rubber Reservoir. Diabetes Care 1 March 1980; 3 (2): 322–331. https://doi.org/10.2337/diacare.3.2.322