We thank Clarke et al. (1) for their thought-provoking response to our article (2). With their comments (1), they not only took on the important issue of how to optimally assess the accuracy of continuous glucose monitors (CGMs); they moved the discussion one step further.

In our study (2), we did indeed take the statistical liberty of deriving degrees of freedom from all pooled data points—in contrast to the proposal by Clarke et al. (1) who compared the accuracy of two sensors using one average mean absolute deviation (MAD) value per patient. The latter approach may be too rigid because not all readings are interdependent. For example, postprandial glucose sensor readings at lunch and at night depend little on each other, if at all. It is common practice to derive degrees of freedom from pooled data in the sensor field. In a previous study, Clarke et al. (3) compared the accuracy of two CGMs in 16 type 1 diabetic patients by using the continuous glucose–error grid analysis (CG-EGA). The difference in pooled readings in the hypoglycemic area that ended up in zones A and B was reported to be highly significant between both sensors (88 vs. 62.8%, respectively) (P < 0.0005). This level of significance implies that degrees of freedom were derived from all data pairs in the hypoglycemic range (250 mg/dl) rather than from the actual amount of participants (n = 16). Even with a strict statistical policy, the better MAD for the microdialysis sensor in the hypoglycemic area in our study (2) (12.0% for the 7-min corrected microdialysis sensor vs. 25.2% for the needle-type sensor, calculated per patient [df = 12], P = 0.036 by Wilcoxon’s signed-rank test) and the larger sensitivity for hypoglycemia associated with this sensor (75.0 [75 data pairs] vs. 55.9% [56 data pairs], P = 0.018 by Pearson’s χ2, with 16 of 16 and 12 of 15 hypoglycemic episodes detected by the microdialysis and needle-type sensor, respectively, P = 0.06 by Pearson’s χ2) contrasted with the CG-EGA that noted no difference (51.5 vs. 60.0% accurate readings and benign errors in the hypoglycemic range [df = 42], P = 0.841 by Pearson’s χ2 for the microdialysis and the needle-type sensor, respectively). Therefore, even with a mild statistical approach (i.e., deriving degrees of freedom from 43 data pairs rather than 13), CG-EGA could not confirm the different accuracy of the sensors in the hypoglycemic range.

As to the order of CGS data points, the sensor’s ability to follow the rate and direction of glucose changes is nicely reflected by the MAD: A sequence of glucose values that has been incorrectly reported by a given sensor (e.g., 90 → 82 → 72 mg/dl instead of 72 → 82 → 90 mg/dl) will result in a worsened MAD.

In reaction to the comment by Clark et al. (1) in regards to time consumption, we were happy to learn that the software for CG-EGA has become available. Nevertheless, the laborious collection of frequent blood samples on fixed intervals (in addition to the construction of a rate, a point accuracy plot, and, finally, a combining matrix) will remain inevitable drawbacks of CG-EGA.

With the attempt to standardize the length of the time intervals, Clark et al. clearly tried to improve the CG-EGA methodology. Nevertheless, a time interval that can vary by 5 min (10–15 min) still leaves the door open for interobserver variability.

As to our finding in a previous study (4) of a 7-min delay that was inherent to the microdialysis instrument itself and not seen in the needle-type sensor, Clarke et al. (1) alluded to a (much-disputed) constant 7-min physiological delay resulting from the relationship between interstitial and blood glucose. This physiological delay has been reported to be anywhere between 0 and 30 min, so the 7-min assumption made for the CG-EGA is questionable. Fortunately, Clarke et al. have now implemented into the software the possibility of setting the delay <7 min.

Currently, the optimal way to assess a CGM seems to be the combination of MAD calculated per glucose range, combined curve fitting with assessment of horizontal and vertical shift, sensitivity, and positive predictive value for detecting hypoglycemia.

1.
Clarke WL, Gonder-Frederick LG, Cox D, Kovatchev B: A critical appraisal of the continuous glucose–error grid analysis (Letter).
Diabetes Care
30
:
449
–450,
2007
2.
Wentholt IM, Hoekstra JB, DeVries JH: A critical appraisal of the continuous glucose–error grid analysis.
Diabetes Care
29
:
1805
–1811,
2006
3.
Clarke WL, Anderson S, Farhy L, Breton M, Gonder-Frederick L, Cox D, Kovatchev B: Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose–error grid analysis.
Diabetes Care
28
:
2412
–2417,
2005
4.
Wentholt IM, Vollebregt MA, Hart AA, Hoekstra JB, DeVries JH: Comparison of a needle-type and a microdialysis continuous glucose monitor in type 1 diabetic patients.
Diabetes Care
28
:
2871
–2876,
2005