OBJECTIVE

The goal was to develop a new index of insulin sensitivity in patients with type 1 diabetes estimated from continuous glucose monitoring (CGM) and subcutaneous insulin delivery data under carefully controlled conditions.

RESEARCH DESIGN AND METHODS

The database consists of 12 subjects with type 1 diabetes, studied during breakfast, lunch, and dinner, in a clinical research unit, wearing both subcutaneous insulin pump and CGM device. Frequent blood samples were drawn for measurements of plasma glucose and insulin concentrations in order to estimate insulin sensitivity with the oral minimal model (SIMM). The new index of insulin sensitivity (SISP) was calculated with a simple algebraic formula for each meal, using only CGM and insulin pump data and compared with SIMM.

RESULTS

SISP was well correlated with SIMM (r = 0.825; P < 10−8), and diurnal pattern was also similar to SIMM.

CONCLUSIONS

A novel method for estimating insulin sensitivity in subjects with type 1 diabetes on sensor-augmented insulin pump therapy has been presented. This new index correlates well with the reference oral minimal model estimate of insulin sensitivity. The knowledge of patient-specific insulin sensitivity and its diurnal variation can help in optimizing insulin therapy in type 1 diabetes and could also inform next-generation closed-loop control systems.

The standard therapy for type 1 diabetes consists of exogenous insulin administration, either by multiple daily injections or with continuous subcutaneous insulin infusion (CSII) through the insulin pump, adjusted according to self-monitored blood glucose (SMBG) levels 3–4 times per day. However, in the past 10–15 years, new possibilities in diabetes therapy have emerged thanks to continuous glucose monitoring (CGM) and CSII, which substitute SMBG and multiple daily injection therapy, respectively. Minimally invasive CGM devices can measure, in real time, interstitial glucose (IG) concentrations in continuous time for up to several days. CSII uses the subcutaneous route and administers insulin with a basal/bolus strategy, i.e., continuously deliver insulin over 24 h and inject boluses.

In order to determine the appropriate meal insulin bolus, it would be important to know the subject-specific insulin sensitivity, i.e., the ability of insulin to stimulate glucose utilization and inhibit glucose production. Indices of insulin sensitivity include those based on hyperinsulinemic-euglycemic clamp (1), intravenous glucose tolerance test (2), and, more recently, meal and oral glucose tolerance test. These indices include the oral glucose minimal model (3,4), a minimal model based integral formula (5), and surrogate measures (68). However, all the above require the measurement of both plasma glucose and insulin concentrations and therefore cannot be used in everyday life. In the outpatient setting, the only usable approach is the risk method proposed by Breton and Kovatchev (9), which uses SMBG data collected over a period of 2–6 weeks and some patient parameters. This index measures the average insulin sensitivity in the preceding 2 weeks and thus cannot be used to assess the intraday variability of insulin sensitivity.

Here we propose a method to estimate insulin sensitivity from CGM sensor and insulin pump data and validate it against the oral minimal model, which uses plasma glucose and insulin concentrations. The method is usable in everyday life in patients wearing the two devices.

Database and Protocol

Twelve type 1 diabetic subjects (5 females, aged 39.5 ± 14.2 years, BMI 25.7 ± 3.8 kg/m2, HbA1c ≤8.5% or 69 mmol/mol) were studied for 3 days in the clinical research unit of the Mayo Clinic Center for Translational Science Activities as part of a separate study to determine diurnal patterns of insulin sensitivity (10). Briefly, once per day, a triple-tracer mixed-meal study protocol was performed during breakfast, lunch, or dinner in Latin square design. Blood samples were collected at −180, −30, 0, 5, 10, 20, 30, 60, 90, 120, 150, 180, 240, 300, and 360, with t = 0 corresponding to meal time, for measurement of plasma glucose and insulin concentrations in order to estimate SI with the oral minimal model (3), here considered as reference. More details can be found in Hinshaw et al. (10). Subjects also wore both subcutaneous insulin pump (Medtronic or Insulet OmniPod) and CGM (Dexcom Seven Plus). Within-subject mean absolute relative difference between CGM readings and reference was equal to 12.1 ± 3.3%, and no systematic variation over time was found (mean [SD] within subjects was 5.3 ± 3.7%). This is perfectly in line with published reports (11). Figure 1 shows CGM and insulin infusion rate in a representative subject. The 12 subjects have been chosen among the 19 reported by Hinshaw et al. (10) since they had the complete CGM and CSII data required for the calculation described below. Of note, one CGM trace was missing in one subject during one meal.

Figure 1

Data needed for the estimation of insulin sensitivity. Glucose data are measured by CGM (left-hand panel, black line), with some SMBG data available for sensor recalibration (left-hand panel, gray dots). Insulin data are the basal infusion and boluses administered by insulin pump (right-hand panel, black line and gray triangles, respectively).

Figure 1

Data needed for the estimation of insulin sensitivity. Glucose data are measured by CGM (left-hand panel, black line), with some SMBG data available for sensor recalibration (left-hand panel, gray dots). Insulin data are the basal infusion and boluses administered by insulin pump (right-hand panel, black line and gray triangles, respectively).

Close modal

Basis

The starting point is the derivation of insulin sensitivity by integrating the oral minimal model equations (3), as described by Caumo et al. (5). Then, the calculation is adapted to allow the use of CGM and CSII data, instead of plasma glucose and insulin concentrations. The oral minimal model (3) is

where G is plasma glucose concentration (milligrams per deciliter), with Gb denoting its basal value; X is insulin action (min−1); I is plasma insulin concentration (microunits per milliliter), with Ib denoting its basal value; RaG is posthepatic appearance of meal glucose (milligrams per kilograms per minute); VG is glucose distribution volume (deciliters per kilogram); p2 is speed of rise and decay of insulin action (min−1); and p3 is its size (min−2 per microunit per milliliter). Insulin sensitivity is defined as

By substituting and into Eq. 1, we obtain

By integrating the differential equation (Eq. 3) from the time of the meal ingestion (tmeal) to the end of the experiment (tend) and rearranging, one has

The first integral in the numerator can be rewritten as

where D (milligrams) is the amount of glucose ingested during the meal and f(tend) is the fraction of the ingested dose, which at t = tend has reached plasma. X′ in Eq. 4 is usually not available because model parameters are unknown, thus the denominator of Eq. 4 is substituted with the average of glucose excursion times the overall insulin stimulus:

where |ΔG| is above basal plasma glucose concentration.

Thus SI is given by

where AUC is the area under the curve calculated from the start of the meal (tmeal) to the end of the experiment (tend). Subject-specific parameters used by the formula are the body weight (BW) (kilograms), height (meters), and age (years). BW is explicitly used in Eq. 7 and, together with height and age, used for the calculation of plasma insulin clearance (CL), as discussed below. Parameter fixed to population values are glucose effectiveness at zero insulin GEZI [deciliters per kilogram per minute; fixed to 0.01 dL/kg/min for diabetic subjects (10,12)] and volume of glucose distribution VG [deciliters per kilogram; fixed to 1.45 dL/kg, according to Dalla Man et al. (3)].

G, ΔG, and I are not directly available but can be derived from CGM and subcutaneous insulin delivery as detailed below.

Glucose Signal From CGM

CGM measures the IG concentration, which is related to plasma glucose by a first-order differential equation, i.e., IG is a delayed version of plasma glucose:

Thus, with a well-calibrated device, assuming IG(tend) = IG(tmeal), one has

i.e.,

Similarly, for the above basal glucose signal, one has

Thus one can safely assume that, in the presence of a well-calibrated device, AUC(ΔCGM) and AUC(|ΔCGM|) are good approximations of AUC(ΔG) and AUC(|ΔG|), respectively.

In presence of a noncalibrated device, if at least two SMBG samples are available, it is possible to offline recalibrate the CGM profile (13,14).

Insulin Signal From CSII

For the calculation of AUC(I), we assume that insulin is not degraded locally in the site of infusion. In this case, the integral of plasma insulin can be obtained from subcutaneous insulin infusion divided by the plasma insulin clearance CL. In fact, plasma insulin kinetics can be described with a single compartment model (15):

where I(t) is the plasma insulin concentration, RaI(t) is the insulin rate of appearance in plasma, VI is the insulin volume of distribution, and n the fractional insulin clearance rate (n = CL/VI).

Then, integrating Eq. 12 from the time of the premeal bolus (tmeal) to the end of the observation period (tend) and assuming that insulin is back to its initial value at the end of the experiment, one has

Finally, under the assumption that all infused insulin eventually reaches the circulation, one has

Thus we can compute AUC(I) from the amount of insulin infused subcutaneously and CL:

where basal(t) is the basal insulin infusion rate during the integration period, bolus(tk) is the premeal or correction bolus administered at t = tk, and CL (liters per minute) is the plasma insulin clearance, which can be calculated from subject’s height, BW, and age by using the population model proposed by Campioni et al. (16). In addition, if correction boluses are administered before the start of the meal, one has to consider that part of that injected insulin could be still active. The residual active insulin can be determined by adopting the same algorithm presented by Patek et al. (17), which uses the insulin on board (IOB) curves adapted from Ellingsen et al. (18). This quantity (IOB[tmeal]) must be added to the previously estimated AUC(I). Moreover, if correction boluses are administered before the end of the considered interval, IOB is used to evaluate the active insulin at the end of the study (IOB[tend]), which is subtracted from the previously estimated AUC(I). In summary,

Accounting for Carbohydrates on Board

As evident from Eq. 7, an accurate calculation of insulin sensitivity requires the knowledge of the amount of carbohydrates entering the circulation in the integration interval [D·f(tend)]. In this study, meals were provided at 7:00 a.m., 1:00 p.m.; and 7:00 p.m. each day, thus the time interval between the meals was at least 6 h. However, this may not always be the case. Furthermore, not all the carbohydrates ingested with a meal may be fully absorbed before the ingestion of a second meal. Thus, to account for unabsorbed carbohydrates, e.g., frequent meals close to each other, the concept of carbohydrates on board (COB) is introduced. Similarly to IOB, COB is a function that, at each time t, quantifies the fraction of the ingested carbohydrates that has not yet appeared in the circulation. COB is based on the model of gastrointestinal tract (19): given the amount of carbohydrates ingested at time tm, COB provides, at each time t > tm, the percentage of carbohydrates not yet absorbed, while for ttm>360 min, it is assumed that the carbohydrate absorption is almost completed. The fraction of the ingested dose that has reached plasma at time t, f(t), can be calculated as the ratio between the AUC of the meal rate of appearance and the ingested dose, D (Fig. 2, left panel), assuming that, at the end of the meal, the fraction of the meal appearing in plasma is f= 0.9 (Fig. 2, right panel). The time course of f is shown in Fig. 2, right panel. COB is then calculated as COB(t) = ff(t).

Figure 2

Time course of meal rate of appearance, RaG (left-hand panel), and fraction of meal that appears in plasma, f (right-hand panel).

Figure 2

Time course of meal rate of appearance, RaG (left-hand panel), and fraction of meal that appears in plasma, f (right-hand panel).

Close modal

To better grasp the use of f(t) and COB(t) for the calculation of insulin sensitivity, let us consider the following example: suppose that a subject eats 50 g of carbohydrates at t = 0 min and another 40 g at t = 180 min then fasts for more than 360 min. One can calculate two values of SI, one for each meal ingestion (SImeal1 and SImeal2). For the calculation of SImeal1, the amount of ingested glucose to be considered is 50·f(180) g of carbohydrates, while for the calculation of the second, SImeal2, one should use 50·COB(180)+40·f.

Generalizing, the amount of carbohydrates to use in the formula for the ith meal is

where AoC is the amount of carbohydrates, D(tmeali) is the amount of carbohydrates ingested at the time of the ith meal (tmeali), tendi is the time at which the ith meal ends (corresponding to the time of ingestion of the [i+1]th meal), and COB(tendi–1) the COB at the end of the (i–1)th meal, which contained D(tmeali–1) amount of carbohydrates.

Insulin Sensitivity From Sensor and Pump Data

Incorporating the above derivations into Eq. 7, one can estimate insulin sensitivity from sensor and pump data (SISP) for the ith meal (20):

It is important to define the domain of validity of Eq. 18. When CGM is still high 6 h after meal ingestion, SISP can become negative. This brings the method working outside its domain of validity. We thus recommend to use the formula only if the recalibrated ΔCGM is lower than 150 mg/dL 6 h after meal ingestion. As a matter of fact, one of our subjects (subject 4) during lunch has recalibrated ΔCGM greater than 150 mg/dL 6 h after meal ingestion, and SISP was negative. This subject was thus excluded in the following analysis.

Minimal Model Insulin Sensitivity and Validation of SISP

The oral glucose minimal model (3,4) was used to estimate insulin sensitivity, from plasma glucose and insulin concentrations, in 12 subjects studied three times (10). Here we consider these measures as reference values (SIMM) to which SISP is compared for validation.

Assessment in Case of Incompletely Absorbed Meals

In our experimental protocol, meals were well spaced (at least 6 h) and thus completely absorbed before the next meal. However, in daily life, this may not happen consistently. Thus it is important to assess the performance of the method in case of incompletely absorbed meals. We did so by comparing estimates of SISP obtained with a 360-min interval to those obtained from shorter intervals (up to 180 min), both with and without using the COB function.

Reproducibility

When proposing a new metric like SISP, it is important to assess its reproducibility, i.e., if it provides similar values when repeatedly applied to the same subject under the same experimental conditions. However, due to the large intrasubject variability of insulin sensitivity (10), SISP will likely change if the same meal is administered in two occasions to a subject. Thus, to address reproducibility, one can resort to simulation. We evaluated SISP reproducibility in simulation using our U.S. Food and Drug Administration-accepted type 1 diabetes simulator (21). We simulated a 7-day scenario for 100 in silico subjects with three meals per day (breakfast from 6:00 to 8:00 a.m., lunch from 11:30 a.m. to 1:30 p.m., and dinner from 6:00 to 8.30 p.m.) with different amounts (breakfast 0.7 – 0.9 g/kg, lunch 0.8 – 1.0 g/kg, and dinner 0.8 – 1.3 g/kg), while subject-specific insulin sensitivity was maintained constant for the whole simulation. A total of 21 SISP were thus calculated for each subject (three values per 7 days). To assess the repeatability of the index, we calculated the average (mean), the SD, and the coefficient of variation (CV; SD/mean) of the 21 estimates of SISP in each subject.

Statistical Analysis

Data are presented as mean ± SD. Two sample comparisons were done by paired sample t test. Pearson’s correlation was used to evaluate univariate correlation.

The correlation between the two indices was very good (r = 0.825; P < 10−8; Fig. 3, right-hand panel), and diurnal pattern was similar, indicating that, apart from a scale factor, SISP closely mirrors SIMM. SIMM and SISP have been estimated in the 12 subjects at breakfast, lunch, and dinner. SISP was significantly higher than SIMM (13.86 ± 14.56 vs. 6.67 ± 5.63 dL/kg/min per μU/mL; P < 10−3; Fig. 3, left-hand panel).

Figure 3

Mean values (left-hand panel) and correlation plot (right-hand panel) between SIMM and SISP insulin sensitivity indices for breakfast, lunch, and dinner (white, striped, and black bars, respectively). *P < 10−3 with paired sample t test.

Figure 3

Mean values (left-hand panel) and correlation plot (right-hand panel) between SIMM and SISP insulin sensitivity indices for breakfast, lunch, and dinner (white, striped, and black bars, respectively). *P < 10−3 with paired sample t test.

Close modal

When SISP was calculated for incompletely absorbed meals, i.e., relying on reduced integration intervals, mean values of SISP were virtually the same (Fig. 4, top left) if one uses the COB function. Conversely, SISP increased systematically if COB is not used (Fig. 4, top right). Of note, the correlation between SISP calculated at the end of the experiment and that obtained from reduced integration intervals decreases only slightly (Fig. 4, middle panels), and the absolute relative error increases (Fig. 4, bottom panels) both with and without using COB. Finally, in silico reproducibility of SISP was 23 ± 6%.

Figure 4

Sensitivity analysis of SISP with (left-hand panel) and without (right-hand panel) accounting for COB for different time integration intervals: mean values of SISP (top), correlation indices (middle), and absolute relative error (bottom), calculated with respect to SISP, i.e., estimated at t = 360 min.

Figure 4

Sensitivity analysis of SISP with (left-hand panel) and without (right-hand panel) accounting for COB for different time integration intervals: mean values of SISP (top), correlation indices (middle), and absolute relative error (bottom), calculated with respect to SISP, i.e., estimated at t = 360 min.

Close modal

Insulin sensitivity is a key parameter of the metabolic status of an individual, which could be beneficial also for optimizing insulin therapy in type 1 diabetes. In fact, the knowledge of patient-specific SI and its daily variation can truly help in determining the optimal insulin bolus to be administered to cover the ingested carbohydrates. However, all methods available for the estimation of insulin sensitivity rely on plasma glucose and insulin measurements and thus cannot be used in everyday life of a patient with type 1 diabetes.

In this article, we have proposed an index of insulin sensitivity, SISP, which can be estimated in patients with type 1 diabetes wearing a CGM sensor and an insulin pump. We have demonstrated that it is similar to the one obtained with the oral minimal model, which requires plasma glucose and insulin data. The method uses retrospective subcutaneous sensor and insulin delivery data with some anthropometric parameters for each subject and provides, for each meal, the patient’s insulin sensitivity by an integral formula.

SISP measures how subcutaneously infused insulin affects the CGM profile. Thus SISP is not exactly the same index derived with the minimal model (SIMM), which represents the ability of insulin to suppress endogenous glucose production and stimulate glucose uptake. In other words, SISP is a new metric of insulin sensitivity with its own range. In fact, mean values of SISP were almost twice SIMM. However, the correlation between the two indices was excellent (r = 0.825; P < 10−8).

A robust estimate of insulin sensitivity can be obtained for each meal whenever meals are well spaced (5–6 h). However, we also tested the method in case of incompletely absorbed meals. To deal with this situation, we introduced the concept of COB. Similarly to IOB (18), COB represents, at each time t, the amount of ingested glucose that has not yet been absorbed. Thus one can define different COB curves for different types of meal, i.e., fast or slow carbohydrates. We used COB in the SISP calculation to evaluate the correct amount of carbohydrates in relation with the observed CGM profiles in a given time interval. We demonstrated the need of using COB; in case of a short time interval between consecutive meals, a robust estimation of SISP can only be obtained if carbohydrates absorption is taken into account (Fig. 4, top left-hand versus top right-hand panel).

Possible applications of the new index include its use for assessing intraday and interday variability (e.g., existence of diurnal patterns) of insulin sensitivity in large cohorts of subjects with type 1 diabetes in normal life conditions. For instance, the method is currently being applied to the Sensor-Augmented Pump Therapy for A1C Reduction (STAR) 3 data of type 1 diabetic subjects wearing a sensor-augmented insulin pump (22). This will provide, in each subject, the SISP time course during several months and will allow testing for the existence of subject-specific SISP daily patterns and correlation of such variation with lifestyle and other factors. Following validation using STAR 3 data, a clinical application would be the use of SISP to calculate the optimal insulin-to-carbohydrate ratio (CR) (23,24). Thus optimizing CR based on recent CGM and CSII data collected 1–2 weeks prior to a clinical visit may be useful for physicians to improve patient-specific meal bolus insulin therapy and thus a major component of glucose control. In addition, the knowledge of patient CR daily pattern may help in the design of optimal closed-loop control algorithms relying on patient-specific open-loop insulin therapy at the time of transition from open- to closed-loop therapy. A natural progression of this work would be the development of real-time adaptation of open- or closed-loop therapy based on the presence of well-developed metrics such as the one developed here. Further development of diabetes technology hardware and algorithms should make this a reality over the next several years.

The method relies on data provided by a CGM device that can occasionally suffer from inaccuracies. For instance, the assumption that AUC(CGM) is a good approximation of AUC(G) becomes critical if the device is not well calibrated. If this occurs, SISP will also reflect sensor inaccuracy. However, to improve the quality of CGM measurements, some algorithms can be used to recalibrate CGM traces (13,14). Another possible limitation is the need to fix some parameters (GEZI and VG) to population values (3,10,12) and others calculated from population models (CL) using anthropometric data (16). In order to test the effect of fixing these parameters, we also calculated SISP using individualized GEZI, estimated with the oral glucose minimal model (3) and CL, directly estimated from the data in each patient; we obtained values very similar to those obtained with fixed parameters and a slightly higher correlation with SIMM.

Insulin sensitivity is an important element in the daily life of patients with type 1 diabetes and could be useful to optimize insulin therapy. However, methods to estimate this index by emerging technologies, such as subcutaneous CGM sensor and insulin pump, has never been proposed. We have presented a method that estimates insulin sensitivity from CGM and insulin pump data. Future studies involving a larger databases that include larger cohorts of subjects studied for a longer time are needed to better define the applicability in free-living conditions.

Acknowledgments. The algorithm proposed in this article is part of a patent request deposited by the University of Padova (13/661,755).

Funding. The work was partially supported by the National Institutes of Health (DK-R01-085561 and DK-DP3-094331); grant number UL1-TR000135 from the National Center for Advancing Translational Sciences, a component of the National Institutes of Health; and the Italian Ministero dell’Istruzione, dell’Università e della Ricerca (Progetto FIRB 2009).

Duality of Interest. No potential conflicts of interest relevant to this article were reported.

Author Contributions. M.S., C.D.M., and C.C. developed the method, analyzed the results, and drafted the manuscript. Y.C.K. and A.B. commented on and edited the manuscript. C.C. is the guarantor of this work and, as such, had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.

Prior Presentation. This work was presented in part at the 11th Annual Diabetes Technology Meeting, San Francisco, CA, 27–29 October 2011, and the 12th Annual Diabetes Technology Meeting, Bethesda, MD, 8–10 November 2012.

1.
DeFronzo
RA
,
Tobin
JD
,
Andres
R
.
Glucose clamp technique: a method for quantifying insulin secretion and resistance
.
Am J Physiol
1979
;
237
:
E214
E223
[PubMed]
2.
Bergman
RN
,
Ider
YZ
,
Bowden
CR
,
Cobelli
C
.
Quantitative estimation of insulin sensitivity
.
Am J Physiol
1979
;
236
:
E667
E677
[PubMed]
3.
Dalla Man
C
,
Caumo
A
,
Basu
R
,
Rizza
R
,
Toffolo
G
,
Cobelli
C
.
Minimal model estimation of glucose absorption and insulin sensitivity from oral test: validation with a tracer method
.
Am J Physiol Endocrinol Metab
2004
;
287
:
E637
E643
[PubMed]
4.
Dalla Man
C
,
Yarasheski
KE
,
Caumo
A
, et al
.
Insulin sensitivity by oral glucose minimal models: validation against clamp
.
Am J Physiol Endocrinol Metab
2005
;
289
:
E954
E959
[PubMed]
5.
Caumo
A
,
Bergman
RN
,
Cobelli
C
.
Insulin sensitivity from meal tolerance tests in normal subjects: a minimal model index
.
J Clin Endocrinol Metab
2000
;
85
:
4396
4402
[PubMed]
6.
Stumvoll
M
,
Mitrakou
A
,
Pimenta
W
, et al
.
Use of the oral glucose tolerance test to assess insulin release and insulin sensitivity
.
Diabetes Care
2000
;
23
:
295
301
[PubMed]
7.
Matsuda
M
,
DeFronzo
RA
.
Insulin sensitivity indices obtained from oral glucose tolerance testing: comparison with the euglycemic insulin clamp
.
Diabetes Care
1999
;
22
:
1462
1470
[PubMed]
8.
Hansen
T
,
Drivsholm
T
,
Urhammer
SA
, et al
.
The BIGTT test: a novel test for simultaneous measurement of pancreatic β-cell function, insulin sensitivity, and glucose tolerance
.
Diabetes Care
2007
;
30
:
257
262
[PubMed]
9.
Breton MD, Kovatchev BP. Method, system and computer program product for evaluation of insulin sensitivity, insulin/carbohydrate ratio, and insulin correction factors in diabetes from self-monitoring data. PCT international patent application PCT/US2008/069416. 8 July 2008
10.
Hinshaw L, Dalla Man C, Nandy DK, et al. Diurnal pattern of insulin action in type 1 diabetes: implications for a closed-loop system. Diabetes 2013;62:2223–2229
11.
Leelarathna
L
,
Nodale
M
,
Allen
JM
, et al
.
Evaluating the accuracy and large inaccuracy of two continuous glucose monitoring systems
.
Diabetes Technol Ther
2013
;
15
:
143
149
[PubMed]
12.
Basu
A
,
Dalla Man
C
,
Basu
R
,
Toffolo
G
,
Cobelli
C
,
Rizza
RA
.
Effects of type 2 diabetes on insulin secretion, insulin action, glucose effectiveness, and postprandial glucose metabolism
.
Diabetes Care
2009
;
32
:
866
872
[PubMed]
13.
Guerra
S
,
Facchinetti
A
,
Sparacino
G
,
Nicolao
GD
,
Cobelli
C
.
Enhancing the accuracy of subcutaneous glucose sensors: a real-time deconvolution-based approach
.
IEEE Trans Biomed Eng
2012
;
59
:
1658
1669
14.
Del Favero
S
,
Facchinetti
A
,
Sparacino
G
,
Cobelli
C
.
Retrofitting algorithm: fusing CGM and references to reconstruct a reliable continuous glucose profile
.
Paris
,
Advanced Technologies and Treatments for Diabetes
,
2013
[PubMed]
15.
Toffolo
GM
,
Campioni
M
,
Basu
R
,
Rizza
RA
,
Cobelli
C
.
A minimal model of insulin secretion and kinetics to assess hepatic insulin extraction
.
Am J Physiol Endocrinol Metab
2006
;
290
:
E169
E176
[PubMed]
16.
Campioni
M
,
Toffolo
G
,
Basu
R
,
Rizza
RA
,
Cobelli
C
.
Minimal model assessment of hepatic insulin extraction during an oral test from standard insulin kinetic parameters
.
Am J Physiol Endocrinol Metab
2009
;
297
:
E941
E948
[PubMed]
17.
Patek
SD
,
Magni
L
,
Dassau
E
, et al
International Artificial Pancreas (iAP) Study Group
.
Modular closed-loop control of diabetes
.
IEEE Trans Biomed Eng
2012
;
59
:
2986
2999
[PubMed]
18.
Ellingsen
C
,
Dassau
E
,
Zisser
H
, et al
.
Safety constraints in an artificial pancreatic beta cell: an implementation of model predictive control with insulin on board
.
J Diabetes Sci Tech
2009
;
3
:
536
544
[PubMed]
19.
Dalla Man
C
,
Camilleri
M
,
Cobelli
C
.
A system model of oral glucose absorption: validation on gold standard data
.
IEEE Trans Biomed Eng
2006
;
53
:
2472
2478
[PubMed]
20.
Cobelli C, Dalla Man C, Schiavon M, Basu A, Kudva Y. Estimation of insulin sensitivity from CGM and subcutaneous insulin delivery in type 1 diabetes. U.S. patent application US 13/661,755. 26 October 2012
21.
Kovatchev
BP
,
Breton
M
,
Man
CD
,
Cobelli
C
.
In silico preclinical trials: a proof of concept in closed-loop control of type 1 diabetes
.
J Diabetes Sci Tech
2009
;
3
:
44
55
[PubMed]
22.
Davis
SN
,
Horton
ES
,
Battelino
T
,
Rubin
RR
,
Schulman
KA
,
Tamborlane
WV
.
STAR 3 randomized controlled trial to compare sensor-augmented insulin pump therapy with multiple daily injections in the treatment of type 1 diabetes: research design, methods, and baseline characteristics of enrolled subjects
.
Diabetes Technol Ther
2010
;
12
:
249
255
[PubMed]
23.
Yodfat O, Shapira G, Gescheit I. A method system and device for assessing insulin sensitivity. PCT publication US20080319381. 31 December 2008
24.
Bevier
WC
,
Zisser
H
,
Palerm
CC
, et al
.
Calculating the insulin to carbohydrate ratio using the hyperinsulinaemic-euglycaemic clamp-a novel use for a proven technique
.
Diabetes Metab Res Rev
2007
;
23
:
472
478
[PubMed]
Readers may use this article as long as the work is properly cited, the use is educational and not for profit, and the work is not altered. See http://creativecommons.org/licenses/by-nc-nd/3.0/ for details.