We studied the accuracy of three techniques for the rapid estimation of the blood glucose concentration with reagent strips in samples of blood from inpatients with diabetes mellitus. We observed that the visual inspection of Dextrostix gave results that were too inaccurate for this technique to be of practical clinical benefit. In contrast, when performed by registered nurses, both the Dextrostix-Eyetone method and the visual inspection of Chemstrips yielded results that were closely correlated with the measured serum glucose concentrations. With both of the latter techniques, less than 11% of the determinations varied from the actual serum glucose concentration by more than 20%. We conclude that both these techniques are sufficiently accurate to be of potential use in the home monitoring of blood glucose concentrations.
Skip Nav Destination
Article navigation
Symposium on Home Glucose Self-Monitoring|
May 01 1981
Comparison of Various Methods for Rapid Glucose Estimation Free
In Special Collection:
Review Articles
Rex S Clements, Jr;
Rex S Clements, Jr
Departments of Medicine and Biostatistics, the School of Nursing, and the Diabetes Research and Training Center of the University of Alabama in Birmingham
Birmingham, Alabama
Search for other works by this author on:
Nancy A Keane;
Nancy A Keane
Departments of Medicine and Biostatistics, the School of Nursing, and the Diabetes Research and Training Center of the University of Alabama in Birmingham
Birmingham, Alabama
Search for other works by this author on:
Katharine A Kirk;
Katharine A Kirk
Departments of Medicine and Biostatistics, the School of Nursing, and the Diabetes Research and Training Center of the University of Alabama in Birmingham
Birmingham, Alabama
Search for other works by this author on:
Buris R Boshell
Buris R Boshell
Departments of Medicine and Biostatistics, the School of Nursing, and the Diabetes Research and Training Center of the University of Alabama in Birmingham
Birmingham, Alabama
Search for other works by this author on:
Address reprint requests to Rex S. Clements, Jr., Veterans Administration Medical Center, 700 South 19th Street, Birmingham, Alabama 35233.
Citation
Rex S Clements, Nancy A Keane, Katharine A Kirk, Buris R Boshell; Comparison of Various Methods for Rapid Glucose Estimation. Diabetes Care 1 May 1981; 4 (3): 392–395. https://doi.org/10.2337/diacare.4.3.392
Download citation file: