Increasing evidence demonstrates the benefits of new diabetes technologies, including insulin pumps and continuous glucose monitors (CGM), for glycemic management in people with type 1 diabetes (T1D). In addition to the independent use of these technologies, hybrid closed-loop systems (HCLS), which combine insulin pumps and CGM with a closed-loop algorithm controller to automate insulin delivery, can improve glucose levels (1,2). This study compared glycemic outcomes in users of HCLS with those of users of insulin pumps and CGM without automated insulin delivery and those using multiple daily insulin injections (MDI) with CGM in youth and adults with T1D.
We analyzed electronic medical records data (2019–2021) from the T1D Exchange Quality Improvement Collaborative (T1DX-QIC), a multicenter database for people with T1D (3). A total of 28,019 people, aged ≥6 years with T1D diagnosis for at least 1 year, were classified into three groups by mode of insulin treatment and CGM use. At their most recent visit, patients who reported using HCLS (either Tandem t:slim X2 pump with Control-IQ or Medtronic 670G or 770G pump with active automated mode) were classified as HCLS users (N = 2,047), those using an insulin pump together with a CGM without automated insulin delivery were classified as Pump+CGM users (N = 12,306), and those using MDI for insulin therapy along with a CGM device were classified as MDI+CGM users (N = 13,613). DIY loopers (patients having built their own closed-loop systems) were excluded from this analysis. Primary outcome was the most recently recorded HbA1c (in percent). Secondary outcomes, available for a subgroup of this population, included time in range (TIR), defined as percentage of time spent between 70 and 180 mg/dL, time below range (TBR) (<70 mg/dL), and time above range (TAR) (>250 mg/dL), using an average of the last 14 days. Data collection for this analysis was approved by the Western Institutional Review Board.
In this study, 58% of the HCLS group, 60% of the Pump+CGM group, and 59% of the MDI+CGM group were individuals ≤18 years of age. Inequities in HCLS use by race/ethnicity and insurance status were observed. Among all non-Hispanic (NH) Black people with T1D, 3% used HCLS, whereas in the NH White population, 8% used HCLS (P < 0.001). Differences in HCLS uptake persisted by race/ethnicity after stratifying by insurance status. Duration of diabetes was longer in the HCLS group than in the Pump+CGM and MDI+CGM groups (mean [SD] 11 [11] vs. 9 [9] and 9 [10] years). In assessment of glycemic outcomes, mean HbA1c levels were lower for HCLS users than for the Pump+CGM group (difference in means [95% CI] −0.5 [−0.6, −0.5]) and the MDI+CGM group (difference in means [95% CI] −0.8 [−0.9, −0.7]; P < 0.001). HbA1c levels were lowest in the HCLS group for both pediatric (aged ≤18 years) and adult (aged >18 years) populations (pediatric median HbA1c [interquartile range] 7.5 [1.6] vs. 8.0 [2.0] and 8.2 [2.4], respectively; adult median HbA1c [interquartile range] 7.0 [1.2] vs. 7.6 [1.8] and 7.7 [2.1], respectively). Linear mixed models, controlling for potential confounding effects of age, sex, race/ethnicity, and insurance status, showed that estimated marginal means (EMM) for HbA1c remained lower among HCLS users than for the Pump+CGM and MDI+CGM comparator groups (EMM [95% CI] 8.1 [8.0, 8.2] vs. 8.7 [8.6, 8.8] and 8.8 [8.7, 8.9]; P < 0.001). Further, among all HCLS users, people of NH Black and Hispanic race/ethnicity had significantly higher adjusted HbA1c levels than NH White peers (EMM [95% CI] 8.4 [7.4, 9.5] and 7.8 [6.8, 9.0] vs. 7.4 [6.4, 8.5]) (Fig. 1A).
Mean TIR was higher in the HCLS group than in the Pump+CGM and MDI+CGM groups (mean [SD] 60% [17] vs. 52% [20] and 50% [21], respectively; P < 0.001), TBR was lower in the HCLS group (mean [SD] 1.6% [1.4] vs. 2.0% [2.3] and 2.1% [2.3], respectively; P < 0.001), and TAR was also lower in the HCLS group (mean [SD] 14% [14] vs. 22% [17] and 21% [19], respectively; P < 0.001). Improved glycemic targets, including more TIR, less TBR, and less TAR, persisted for the HCLS group after adjustment for the abovementioned covariates for the overall population (Fig. 1B) as well as for the pediatric and adult populations separately.
This report demonstrates the value of HCLS use in lowering glycemic outcomes; however, a limitation of this cross-sectional study is that we were unable to rule out reverse causation, in that people with lower HbA1c may be more likely to adopt HCLS. While the benefit of HCLS technology is apparent for both children and adults, the adult population showed better glycemic levels than the pediatric group, potentially owing to the inherent challenges in reaching optimal glycemic targets in children and young adults with T1D (4). Lastly, advanced diabetes technology use was lower in NH Black and publicly insured people, indicating that social disparities continue to be a hindrance to better health outcomes in this population (5). There remains a need to address inequities for vulnerable groups with diabetes.
O.E. and D.J.D. are co-senior authors.
Article Information
Acknowledgments. The authors thank all team members and patients across all participating centers in the T1DX-QIC.
Funding. The T1DX-QIC is supported by the The Leona M. and Harry B. Helmsley Charitable Trust.
Duality of Interest. O.E. is a member of the Medtronic Diabetes Health Inequity Board. D.J.D. has served as an independent consultant for Dexcom and Insulet, and his institution has received research support from Insulet separate from this work.
Author Contributions. N.N. and O.E. conceptualized the study, N.N. wrote the draft manuscript and analyzed the data, and N.N., O.E., and D.J.D. contributed to the interpretation of data. All authors critically reviewed the manuscript and approved the final version. N.N. is the guarantor of this work and, as such, had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.