We designed and constructed a new miniature, open-loop insulin infusion pump specifically to overcome the problems of many first generation insulin infusers. Special features are small size, adjustable volumetric basal infusion rate, rapid electronically mediated prandial insulin boosts, facility for doubling and halving the basal infusion rate and/or prandial delivery, and alarms for low battery state, motor over-run, stoppage, and control circuit malfunction. The infuser takes a specially designed syringe prefilled with short-acting insulin, sufficient in most diabetic patients for at least 7 days treatment with 100 U/ml insulin. To test clinical efficacy nine insulin-dependent diabetic patients received continuous subcutaneous insulin infusion (CSII) with the new infuser for periods up to 6 mo. Four patients previously CSII-treated with a first-generation pump and five who were new to CSII achieved and maintained the expected degree of near-normoglycemia. There were no pump breakdowns and a questionnaire completed by patients during the study confirmed ease and simplicity of operation and an appreciation of the advantages of the new pump compared with one widely used first-generation infuser.
Skip Nav Destination
Article navigation
Original Articles|
September 01 1983
Meeting the Problems of First-Generation Insulin Infusion Pumps: Clinical Trial of a New Miniature Infuser
Jeremy J Bending, MRCP;
Jeremy J Bending, MRCP
Unit for Metabolic Medicine and the Department of Chemical Pathology, Guy's Hospital Medical School
London, United Kingdom
National Institute for Medical Research
Mill Hill, London, United Kingdom
Search for other works by this author on:
John C Pickup, DPhil;
John C Pickup, DPhil
Unit for Metabolic Medicine and the Department of Chemical Pathology, Guy's Hospital Medical School
London, United Kingdom
National Institute for Medical Research
Mill Hill, London, United Kingdom
Search for other works by this author on:
Harry Keen, MD;
Harry Keen, MD
Unit for Metabolic Medicine and the Department of Chemical Pathology, Guy's Hospital Medical School
London, United Kingdom
National Institute for Medical Research
Mill Hill, London, United Kingdom
Search for other works by this author on:
Denis Rothwell, MA;
Denis Rothwell, MA
Unit for Metabolic Medicine and the Department of Chemical Pathology, Guy's Hospital Medical School
London, United Kingdom
National Institute for Medical Research
Mill Hill, London, United Kingdom
Search for other works by this author on:
Ian A Sutherland, PhD
Ian A Sutherland, PhD
Unit for Metabolic Medicine and the Department of Chemical Pathology, Guy's Hospital Medical School
London, United Kingdom
National Institute for Medical Research
Mill Hill, London, United Kingdom
Search for other works by this author on:
Address reprint requests to Dr. J. J. Bending, Unit for Metabolic Medicine, Guy's Hospital Medical School, London SE1 9RT, United Kingdom
Citation
Jeremy J Bending, John C Pickup, Harry Keen, Denis Rothwell, Ian A Sutherland; Meeting the Problems of First-Generation Insulin Infusion Pumps: Clinical Trial of a New Miniature Infuser. Diabetes Care 1 September 1983; 6 (5): 452–458. https://doi.org/10.2337/diacare.6.5.452
Download citation file: