A new double-tipped strip for home glucose monitoring without a meter is described. Comparison between values obtained with the new method by trained nursing staff and those obtained with a glucose-oxidase analyzer demonstrates close correlation between the two methods (r = +0.95, a = +22.2 mg/dl, b = +1.14). The mean concentration in 117 strip tests over an analyzer range of 5–775 mg/dl (0.3–43.0 mmol/L) was 52 mg/dl (2.9 mmol/L) greater than the analyzer estimate. Forty-seven percent of samples were read as the color block closest to the analyzer estimate. Samples in the lower glucose range tended to read one block high. The technique is faster to perform than most other visual blood glucose strip methods, and is a useful addition to the equipment available for home glucose monitoring. Modifications to the strip by the manufacturers appear to have overcome the low value over-reading error.
Skip Nav Destination
Article navigation
Original Articles|
November 01 1983
Assessment of a New Visual Blood Glucose Strip
Adrian J L Clark, M.R.C.P.;
Adrian J L Clark, M.R.C.P.
Unit for Metabolic Medicine, Department of Medicine, Guy's Hospital Medical School
London
Search for other works by this author on:
Rosemarie D Cudd, S.R.N.;
Rosemarie D Cudd, S.R.N.
Unit for Metabolic Medicine, Department of Medicine, Guy's Hospital Medical School
London
Search for other works by this author on:
Colette Newey, S.R.N.;
Colette Newey, S.R.N.
Unit for Metabolic Medicine, Department of Medicine, Guy's Hospital Medical School
London
Search for other works by this author on:
Deirdre Brooke, S.R.N.;
Deirdre Brooke, S.R.N.
Unit for Metabolic Medicine, Department of Medicine, Guy's Hospital Medical School
London
Search for other works by this author on:
Ronald G Newall, M.Phil;
Ronald G Newall, M.Phil
Miles Laboratories Limited, Stoke Poges
Slough, United Kingdom
Search for other works by this author on:
Harry Keen, F.R.C.P.
Harry Keen, F.R.C.P.
Unit for Metabolic Medicine, Department of Medicine, Guy's Hospital Medical School
London
Search for other works by this author on:
Address reprint requests to Prof. H. Keen, Unit for Metabolic Medicine, Department of Medicine, 4th floor, Hunt's House, Guy's Hospital Medical School, London SE1 9RT, United Kingdom.
1
The results of this study were presented to the British Diabetic Association Medical and Scientific Meeting, Dublin, September 1982.
Citation
Adrian J L Clark, Rosemarie D Cudd, Colette Newey, Deirdre Brooke, Ronald G Newall, Harry Keen; Assessment of a New Visual Blood Glucose Strip. Diabetes Care 1 November 1983; 6 (6): 540–542. https://doi.org/10.2337/diacare.6.6.540
Download citation file: