We discovered that skilled nurses only casually trained in the use of a fingertip blood glucose reflectance meter (Glucoscan, Lifescan, Mountainview, California) had a 36% incidence of unacceptable results (> 15% from reference). A controlled study was undertaken and showed that with Glucoscan I (GI) 4 of 27 readings were unacceptable and with Glucoscan II (GII) 3 of 27 readings were unacceptable, a statistically nonsignificant difference. Minor deviations from the manufacturer's recommended technique had a significant effect on the results with GI. In contrast, GII was much less sensitive to variations in recommended technique. GI underestimated the reference glucose concentration by 11.7%, and GII overestimated by 6.5%, a statistically significant difference. We conclude that the health professional must be aware of interdevice and intradevice variability in self-monitoring of blood glucose (SMBG). Patients need careful training in the method of SMBG. The results of any single value should be interpreted with caution.
Skip Nav Destination
Article navigation
Technical Section|
November 01 1984
Potential Pitfalls in the Use of Glucoscan and Glucoscan II Meters for Self-Monitoring of Blood Glucose
Victor G Laus, M.S.;
Victor G Laus, M.S.
Timken Mercy Medical Center and Northeastern Ohio Universities College of Medicine
Canton, Ohio
Search for other works by this author on:
Mark A Dietz, B.S.;
Mark A Dietz, B.S.
Timken Mercy Medical Center and Northeastern Ohio Universities College of Medicine
Canton, Ohio
Search for other works by this author on:
Richard P Levy, M.D.
Richard P Levy, M.D.
Timken Mercy Medical Center and Northeastern Ohio Universities College of Medicine
Canton, Ohio
Search for other works by this author on:
Address reprint requests to Richard P. Levy, M.D., Timken Mercy Medical Center, 1320 Timken Mercy Drive N.W., Canton, Ohio 44708.
1
This study was presented in part at the 43d Annual Meeting of the American Diabetes Association, San Antonio, Texas, 1983.
Citation
Victor G Laus, Mark A Dietz, Richard P Levy; Potential Pitfalls in the Use of Glucoscan and Glucoscan II Meters for Self-Monitoring of Blood Glucose. Diabetes Care 1 November 1984; 7 (6): 590–594. https://doi.org/10.2337/diacare.7.6.590
Download citation file: