The t:slim X2 insulin pump with Control-IQ technology (Control-IQ) advanced hybrid closed-loop automated insulin delivery system was evaluated in this prospective single-arm trial. Thirty adults with type 2 diabetes using the Control-IQ system showed substantial glycemic improvement with no increase in hypoglycemia. Mean time in range (70–180 mg/dL) improved 15%, representing an increase of 3.6 hours/day, and mean glucose decreased by 22 mg/dL.

The benefits of hybrid closed-loop automated insulin delivery (AID) systems are well established in type 1 diabetes, with three currently commercially available systems approved by the U.S. Food and Drug Administration (FDA) (17). However, the benefits of AID technology in individuals with type 2 diabetes has not been well established in clinical trials, and none of the systems are currently FDA-approved for use in type 2 diabetes.

We conducted a single-arm, prospective, multicenter study to provide preliminary data on the safety and efficacy of using the t:slim X2 insulin pump with Control-IQ technology (Control-IQ; Tandem Diabetes Care, San Diego, CA) in 30 adults with type 2 diabetes previously using a multiple daily injection (MDI) insulin regimen, insulin pump therapy, or a basal-only insulin regimen.

The study was conducted at three diabetes centers in the United States. The protocol was approved by a central institutional review board, and informed consent was obtained from each participant. The FDA approved an investigational device exemption for the conduct of the trial. The trial is registered at ClinicalTrials.gov (NCT05111301), and a full list of study site personnel is provided (Supplementary Material).

To be eligible for the trial, potential participants had to have type 2 diabetes treated with either basal-bolus insulin therapy (MDI or via an insulin pump) or basal-only insulin (without bolus insulin) for at least 3 months, have a total daily insulin dose (TDD) ≤200 units/day, and have an A1C of 7.5–12.0% (58–108 mmol/mol). Glucose-lowering medications in addition to insulin were permitted provided they were started >3 months before enrollment and their dose was stable. A complete list of inclusion and exclusion criteria is provided in Supplementary Table S1. A list of noninsulin glucose- lowering medications and dosages for each participant is provided in Supplementary Table S2.

Eligible participants initiated use of an unblinded continuous glucose monitoring (CGM) system with their existing insulin therapy and adjunctive therapies for 2–4 weeks. Participants using a Dexcom G6 CGM system with ≥85% of possible glucose values captured during the 14 days before enrollment could skip the CGM run-in period. A successful CGM run-in (≥85% use) was followed by open-loop use of the study pump (i.e., without Control-IQ AID functionality operational) and CGM for 2–4 weeks. Insulin therapy adjustments were made to optimize open-loop glycemic control.

Participants who successfully completed the open-loop pump run-in (defined as using the pump for the specified period with infusion set changes, meal bolus announcements, and no safety concerns) were trained on the use of the study pump in closed-loop mode (i.e., with Control-IQ functionality activated). Instructions were provided for management of hypoglycemia, prolonged hyperglycemia, and sick days/illnesses. Participants were asked to use the system’s programmed sleep mode each night and asked to perform an exercise session with the exercise mode enabled for ≥30 minutes at least once per week. In addition to the study CGM system, AID pump, and related supplies, participants were provided with a blood glucose meter (Contour Next One, Ascensia Diabetes Care, Parsippany, NJ) and a blood ketone meter (Precision Xtra, Abbott Diabetes Care, Alameda, CA). After training, the 6-week period of Control-IQ use was initiated.

Phone contacts or clinic visits occurred 3 days (phone), 1 week (phone), 2 weeks (clinic), 3 weeks (phone), and 6 weeks (clinic) after initiation of Control-IQ. The occurrence of device issues or adverse events were solicited at each visit and contact. Weight and height were measured at baseline, with a follow-up weight measured at 6 weeks. A1C and random C-peptide (Cobas e 801; Roche Diagnostics, Basel, Switzerland) were measured at baseline at a central laboratory. Patient- reported outcome (PRO) questionnaires were completed at screening, at the start of Control-IQ use, and at 6 weeks. Details about testing at each visit are provided in Supplementary Table S3.

Safety outcomes included severe hypoglycemia (defined as hypoglycemia requiring assistance because of an altered cognitive state), diabetic ketoacidosis (DKA; defined by the criteria established in the Diabetes Control and Complications Trial) (8), hyperosmolar hyperglycemia syndrome (HHS), and other adverse events. CGM outcomes measured over 6 weeks included time in range (TIR; 70–180 mg/dL), time >180 mg/dL, mean glucose, time <54 mg/dL, and additional hyperglycemia and hypoglycemia metrics. TDD and body weight were also assessed. PRO measures included the DAWN Impact of Diabetes Profile (DIDP), the Diabetes Impact and Device Satisfaction (DIDS) measure, the PROMIS Sleep-Related Impairment Questionnaire, and the System Usability Scale. A detailed list of all questionnaires administered in this study is provided in Supplementary Table S4.

Analyses were conducted overall and separately for basal-bolus insulin users, basal-only insulin users, and users of sodium–glucose cotransporter 2 (SGLT2) inhibitor or glucagon-like peptide 1 (GLP-1) receptor agonist medications. Baseline CGM metrics were calculated from the last 7 days of the CGM run-in period and 14 days of the open-loop insulin pump period. Follow-up metrics were calculated over the 6 weeks of Control-IQ use. Analyses were conducted overall, for daytime (6:00 a.m. to 11:59 p.m.) and nighttime (midnight to 5:59 a.m.), and for the periods during and after each at-home exercise session.

Summary statistics appropriate to the distribution (mean ± SD or median [interquartile range (IQR)]) were calculated for each CGM metric at baseline, during Control-IQ use, and for change. CGM metrics during Control-IQ use were compared with baseline values (obtained during the unblinded CGM run-in period) using paired t tests when the distribution was normally distributed and a robust regression when the distribution was skewed. The false discovery rate was used to adjust for multiple comparisons (9). Analyses were conducted using SAS, v. 9.4, statistical software (SAS Institute, Cary, NC).

Of the 42 adults who were enrolled into screening, 10 did not meet eligibility criteria and 2 were withdrawn by the site because of nonadherence during the pump run-in phase. Among the 30 who initiated use of Control-IQ, the mean age was 54 ± 12 years, mean A1C was 8.6 ± 1.2% (70 ± 13.1 mmol/mol), and mean BMI was 31 ± 6 kg/m2. Prior insulin delivery was with an MDI regimen in 15 participants, an insulin pump in 2, and basal insulin only in 13. Metformin was being used by 21 (70%), and 22 (73%) were using an SGLT2 inhibitor or GLP-1 receptor agonist (Table 1).

TABLE 1

Participant Characteristics

Overall (n = 30)MDI/Pump (n = 17)Basal-Only (n = 13)
Age, years 54 ± 12 54 ± 12 54 ± 12 
Female sex 18 (60) 8 (47) 10 (77) 
Race/ethnicity
 White non-Hispanic
 White Hispanic or Latino
 Black
 Asian
 American Indian/Alaskan Native
 Unknown/not reported 

17 (57)
6 (20)
4 (13)
1 (3)
1 (3)
1 (3) 

10 (59)
4 (24)
1 (6)
1 (6)
0 (0)
1 (6) 

7 (54)
2 (15)
3 (23)
0 (0)
1 (8)
0 (0) 
Education
 Less than bachelor’s degree
 Bachelor’s or associate degree
 Greater than bachelor’s degree 

8 (27)
16 (53)
6 (20) 

4 (24)
10 (59)
3 (18) 

4 (31)
6 (46)
3 (23) 
Annual household income, $
 <50,000
 50,000–<100,000
 100,000–<200,000
 ≥200,000
 Did not provide/unknown 

4 (13)
5 (17)
12 (40)
4 (13)
5 (17) 

2 (12)
3 (18)
6 (35)
2 (12)
2 (24) 

2 (15)
2 (15)
6 (46)
2 (15)
1 (8) 
Diabetes duration, years 14 (10–22) 13 (10–24) 15 (7–19) 
BMI, kg/m2 31 ± 6 33 ± 7 30 ± 5 
A1C, % 8.6 ± 1.2 8.9 ± 1.3 8.2 ± 0.9 
C-peptide, nmol/L range 0.8 (0.5–1.3)
0.1–2.4 
0.7 (0.5–1.0)
0.1–1.4 
0.9 (0.5–1.4)
0.3–2.4 
CGM use
 Current*
 In the past
 Never 

16 (53)
5 (17)
9 (30) 

11 (65)
2 (12)
4 (24) 

5 (38)
3 (23)
5 (38) 
Insulin modality
 Basal insulin only
 MDI regimen
 Insulin pump 

13 (43)
15 (50)
2 (7) 


15 (88)
2 (12) 

13 (100)

— 
Number and types of noninsulin glucose-lowering medications
 0 Medications
 1 Medication
  Metformin only
  GLP-1 receptor agonist only
  SGLT2 inhibitor only
 2 Medications
  Metformin and GLP-1 receptor agonist
  Metformin and SGLT2 inhibitor
  Metformin and dipeptidyl peptidase 4 inhibitor
  GLP-1 receptor agonist and SGLT2 inhibitor
 3 Medications (metformin, GLP-1 receptor agonist, and SGLT2 inhibitor)
 4 Medications (metformin, GLP-1 receptor agonist, SGLT2 inhibitor, and pioglitazone) 

1 (3)
13 (43)
6 (20)
6 (20)
1 (3)
14 (47)
6 (20)
6 (20)
1 (3)
1 (3)
1 (3)
1 (3) 

1 (6)
8 (47)
4 (24)
4 (24)
0 (0)
7 (41)
4 (24)
3 (18)
0 (0)
0 (0)
1 (6)
0 (0) 

0 (0)
5 (38)
2 (15)
2 (15)
1 (8)
7 (54)
2 (15)
3 (23)
1 (8)
1 (8)
0 (0)
1 (8) 
No severe hypoglycemia episodes during the past 12 months 30 (100) 17 (100) 13 (100) 
No DKA episodes during the past 12 months 30 (100) 17 (100) 13 (100) 
Overall (n = 30)MDI/Pump (n = 17)Basal-Only (n = 13)
Age, years 54 ± 12 54 ± 12 54 ± 12 
Female sex 18 (60) 8 (47) 10 (77) 
Race/ethnicity
 White non-Hispanic
 White Hispanic or Latino
 Black
 Asian
 American Indian/Alaskan Native
 Unknown/not reported 

17 (57)
6 (20)
4 (13)
1 (3)
1 (3)
1 (3) 

10 (59)
4 (24)
1 (6)
1 (6)
0 (0)
1 (6) 

7 (54)
2 (15)
3 (23)
0 (0)
1 (8)
0 (0) 
Education
 Less than bachelor’s degree
 Bachelor’s or associate degree
 Greater than bachelor’s degree 

8 (27)
16 (53)
6 (20) 

4 (24)
10 (59)
3 (18) 

4 (31)
6 (46)
3 (23) 
Annual household income, $
 <50,000
 50,000–<100,000
 100,000–<200,000
 ≥200,000
 Did not provide/unknown 

4 (13)
5 (17)
12 (40)
4 (13)
5 (17) 

2 (12)
3 (18)
6 (35)
2 (12)
2 (24) 

2 (15)
2 (15)
6 (46)
2 (15)
1 (8) 
Diabetes duration, years 14 (10–22) 13 (10–24) 15 (7–19) 
BMI, kg/m2 31 ± 6 33 ± 7 30 ± 5 
A1C, % 8.6 ± 1.2 8.9 ± 1.3 8.2 ± 0.9 
C-peptide, nmol/L range 0.8 (0.5–1.3)
0.1–2.4 
0.7 (0.5–1.0)
0.1–1.4 
0.9 (0.5–1.4)
0.3–2.4 
CGM use
 Current*
 In the past
 Never 

16 (53)
5 (17)
9 (30) 

11 (65)
2 (12)
4 (24) 

5 (38)
3 (23)
5 (38) 
Insulin modality
 Basal insulin only
 MDI regimen
 Insulin pump 

13 (43)
15 (50)
2 (7) 


15 (88)
2 (12) 

13 (100)

— 
Number and types of noninsulin glucose-lowering medications
 0 Medications
 1 Medication
  Metformin only
  GLP-1 receptor agonist only
  SGLT2 inhibitor only
 2 Medications
  Metformin and GLP-1 receptor agonist
  Metformin and SGLT2 inhibitor
  Metformin and dipeptidyl peptidase 4 inhibitor
  GLP-1 receptor agonist and SGLT2 inhibitor
 3 Medications (metformin, GLP-1 receptor agonist, and SGLT2 inhibitor)
 4 Medications (metformin, GLP-1 receptor agonist, SGLT2 inhibitor, and pioglitazone) 

1 (3)
13 (43)
6 (20)
6 (20)
1 (3)
14 (47)
6 (20)
6 (20)
1 (3)
1 (3)
1 (3)
1 (3) 

1 (6)
8 (47)
4 (24)
4 (24)
0 (0)
7 (41)
4 (24)
3 (18)
0 (0)
0 (0)
1 (6)
0 (0) 

0 (0)
5 (38)
2 (15)
2 (15)
1 (8)
7 (54)
2 (15)
3 (23)
1 (8)
1 (8)
0 (0)
1 (8) 
No severe hypoglycemia episodes during the past 12 months 30 (100) 17 (100) 13 (100) 
No DKA episodes during the past 12 months 30 (100) 17 (100) 13 (100) 

Data are mean ± SD, n (%), or median (IQR) unless otherwise noted.

*

Twelve used a FreeStyle Libre and four used a Dexcom G6 CGM system.

Among 13 basal-only insulin users, seven used glargine (1 injection/day), two used glargine (2 injections/day), two used NPH insulin (2 injections/day), one used degludec (1 injection/day), and one used detemir (1 injection/day). Among 15 MDI users, 12 used glargine (1 injection/day), two used degludec (1 injection/day), and one used detemir (1 injection/day) in conjunction with injections of a rapid-acting insulin.

All 30 participants completed the 6-week trial, with 100% of scheduled visits and contacts completed. There were four unscheduled visits and 84 unscheduled contacts among the 30 participants during the 6 weeks of Control-IQ use. Common reasons for these contacts included exercise session logistics (28%), reminders about device time changes for Daylight Saving Time (18%), diabetes management issues (18%), study supplies logistics (13%), device training (9%), and protocol/procedures training (6%).

Control-IQ was active a median of 96% (IQR 91–97%, range 57–99%) of possible time during the 6 weeks, with 23 participants (77%) in closed-loop functionality ≥90% of the time and 27 (90%) in closed-loop functionality ≥80% of the time. The three participants with <80% of time in closed-loop functionality had Control-IQ active for 78%, 71% (reported running out of study CGM supplies), and 57% (reported running out of study insulin) of the 6-week period, respectively.

Mean TIR increased from 56% at baseline to 71% with Control-IQ (mean change 15%, 95% CI 5–24%, P = 0.007), representing an increase of 3.6 hours/day (Table 2 and Figure 1A and B). The beneficial effect was evident over the entire 24 hours of the day (Figure 1C and Supplementary Table S5). Improvement in TIR occurred rapidly, starting with the first day of Control-IQ use, and mean TIR was stable thereafter (Figure 2). The higher the baseline A1C was, the greater the treatment effect was (P = 0.05) (Supplementary Figure S1). In the 17 participants with a baseline A1C ≥8.0% (64 mmol/mol), mean TIR increased by 19 ± 32% compared with 9 ± 13% among the 13 participants with a baseline A1C <8.0% (64 mmol/mol) (Supplementary Figure S2).

TABLE 2

Glycemic Outcomes for Participants Overall, MDI Regimen/Insulin Pump Users, and Basal-Only Insulin Users

BaselineOpen-LoopClosed-LoopMean Estimate (95% CI) for Closed-Loop Versus BaselineP for Closed-Loop Versus Baseline
Overall (n = 30) 
CGM data, hours 164 (150–166) 309 (285–402) 835 (762–883) — — 
TIR (70–180 mg/dL), % 56 ± 25 63 ± 20 71 ± 18 15 (5–24) 0.007 
TIR (70–140 mg/dL), % 30 ± 17 34 ± 22 42 ± 18 13 (5–20) 0.007 
Glucose, mg/dL 184 ± 42 172 ± 32 163 ± 28 −22 (−37 to −6) 0.01 
Time >180 mg/dL, % 44 ± 25 37 ± 20 29 ± 18 −15 (−24 to −5) 0.007 
Time >250 mg/dL, % 8.5 (1.7–21.1) 6.7 (2.1–12.2) 2.9 (1.1–12.8) −3.7 (−7.0 to −0.3) 0.03 
Time >300 mg/dL, % 0.8 (0.0–8.4) 0.6 (0.0–4.8) 0.5 (0.0–3.6) −0.9 (−1.8 to 0.0) 0.05 
Time <70 mg/dL, % 0.05 (0.00–0.47) 0.10 (0.00–0.51) 0.12 (0.04–0.34) −0.03% (−0.13 to 0.07) 0.54 
Time <54 mg/dL, % 0.00 (0.00–0.11) 0.00 (0.00–0.08) 0.02 (0.00–0.07) 0.00 (−0.02 to 0.02) 0.92 
Coefficient of variation, % 28 (24–31) 29 (23–34) 28 (24–31) −0.2 (−2.0 to 1.5) 0.80 
MDI/pump users (n = 17) 
CGM data, hours 162 (151–166) 309 (307–402) 824 (762–892) — — 
TIR (70–180 mg/dL), % 52 ± 31 61 ± 18 68 ± 20 17 (−7 to 40) — 
TIR (70–140 mg/dL), % 28 ± 20 31 ± 21 40 ± 19 12 (−3 to 28) — 
Glucose, mg/dL 192 ± 52 174 ± 30 166 ± 32 −26 (−65 to 13) — 
Time >180 mg/dL, % 48 ± 31 39 ± 19 31 ± 20 −16 (−40 to 7) — 
Time >250 mg/dL, % 6.8 (2.4–29.6) 6.3 (3.8–10.0) 2.9 (1.3–13.1) −4.3 (−12.8 to 4.2) — 
Time >300 mg/dL, % 0.7 (0.0–20.4) 0.5 (0.4–1.6) 0.5 (0.1–4.2) −0.6 (−2.0 to 0.9) — 
Time <70 mg/dL, % 0.11 (0.00–0.47) 0.03 (0.00–0.60) 0.13 (0.03–0.36) −0.06 (−0.30 to 0.18) — 
Time <54 mg/dL, % 0.00 (0.00–0.00) 0.00 (0.00–0.03) 0.02 (0.00–0.05) 0.00 (−0.05 to 0.05) — 
Coefficient of variation, % 25 (23–32) 29 (23–34) 30 (24–31) 1.0 (−2.2 to 4.2) — 
Basal-only insulin users (n = 13) 
CGM data, hours 164 (147–165) 307 (285–382) 845 (770–882) — — 
TIR (70–180 mg/dL), % 61 ± 15 66 ± 22 74 ± 15 13 (3–22) — 
TIR (70–140 mg/dL), % 32 ± 13 38 ± 23 45 ± 17 13 (1–24) — 
Glucose, mg/dL 174 ± 22 170 ± 35 159 ± 23 −16 (−29 to −2) — 
Time >180 mg/dL, % 39 ± 15 34 ± 22 26 ± 15 −12 (−22 to −3) — 
Time >250 mg/dL, % 12.9 (1.7–14.7) 8.1 (0.9–16.9) 1.9 (0.8–7.5) −4.3 (−8.8 to 0.2) — 
Time >300 mg/dL, % 1.6 (0.0–4.9) 0.6 (0.0–5.5) 0.2 (0.0–2.0) −1.1 (−2.2 to 0.1) — 
Time <70 mg/dL, % 0.00 (0.00–0.25) 0.19 (0.00–0.35) 0.09 (0.05–0.25) 0.03 (−0.01 to 0.08) — 
Time <54 mg/dL, % 0.00 (0.00–0.11) 0.00 (0.00–0.12) 0.02 (0.01–0.08) 0.00 (−0.03 to 0.02) — 
Coefficient of variation, % 29 (27–31) 30 (23–31) 27 (24–30) −1.6 (−4.0 to 0.7) — 
BaselineOpen-LoopClosed-LoopMean Estimate (95% CI) for Closed-Loop Versus BaselineP for Closed-Loop Versus Baseline
Overall (n = 30) 
CGM data, hours 164 (150–166) 309 (285–402) 835 (762–883) — — 
TIR (70–180 mg/dL), % 56 ± 25 63 ± 20 71 ± 18 15 (5–24) 0.007 
TIR (70–140 mg/dL), % 30 ± 17 34 ± 22 42 ± 18 13 (5–20) 0.007 
Glucose, mg/dL 184 ± 42 172 ± 32 163 ± 28 −22 (−37 to −6) 0.01 
Time >180 mg/dL, % 44 ± 25 37 ± 20 29 ± 18 −15 (−24 to −5) 0.007 
Time >250 mg/dL, % 8.5 (1.7–21.1) 6.7 (2.1–12.2) 2.9 (1.1–12.8) −3.7 (−7.0 to −0.3) 0.03 
Time >300 mg/dL, % 0.8 (0.0–8.4) 0.6 (0.0–4.8) 0.5 (0.0–3.6) −0.9 (−1.8 to 0.0) 0.05 
Time <70 mg/dL, % 0.05 (0.00–0.47) 0.10 (0.00–0.51) 0.12 (0.04–0.34) −0.03% (−0.13 to 0.07) 0.54 
Time <54 mg/dL, % 0.00 (0.00–0.11) 0.00 (0.00–0.08) 0.02 (0.00–0.07) 0.00 (−0.02 to 0.02) 0.92 
Coefficient of variation, % 28 (24–31) 29 (23–34) 28 (24–31) −0.2 (−2.0 to 1.5) 0.80 
MDI/pump users (n = 17) 
CGM data, hours 162 (151–166) 309 (307–402) 824 (762–892) — — 
TIR (70–180 mg/dL), % 52 ± 31 61 ± 18 68 ± 20 17 (−7 to 40) — 
TIR (70–140 mg/dL), % 28 ± 20 31 ± 21 40 ± 19 12 (−3 to 28) — 
Glucose, mg/dL 192 ± 52 174 ± 30 166 ± 32 −26 (−65 to 13) — 
Time >180 mg/dL, % 48 ± 31 39 ± 19 31 ± 20 −16 (−40 to 7) — 
Time >250 mg/dL, % 6.8 (2.4–29.6) 6.3 (3.8–10.0) 2.9 (1.3–13.1) −4.3 (−12.8 to 4.2) — 
Time >300 mg/dL, % 0.7 (0.0–20.4) 0.5 (0.4–1.6) 0.5 (0.1–4.2) −0.6 (−2.0 to 0.9) — 
Time <70 mg/dL, % 0.11 (0.00–0.47) 0.03 (0.00–0.60) 0.13 (0.03–0.36) −0.06 (−0.30 to 0.18) — 
Time <54 mg/dL, % 0.00 (0.00–0.00) 0.00 (0.00–0.03) 0.02 (0.00–0.05) 0.00 (−0.05 to 0.05) — 
Coefficient of variation, % 25 (23–32) 29 (23–34) 30 (24–31) 1.0 (−2.2 to 4.2) — 
Basal-only insulin users (n = 13) 
CGM data, hours 164 (147–165) 307 (285–382) 845 (770–882) — — 
TIR (70–180 mg/dL), % 61 ± 15 66 ± 22 74 ± 15 13 (3–22) — 
TIR (70–140 mg/dL), % 32 ± 13 38 ± 23 45 ± 17 13 (1–24) — 
Glucose, mg/dL 174 ± 22 170 ± 35 159 ± 23 −16 (−29 to −2) — 
Time >180 mg/dL, % 39 ± 15 34 ± 22 26 ± 15 −12 (−22 to −3) — 
Time >250 mg/dL, % 12.9 (1.7–14.7) 8.1 (0.9–16.9) 1.9 (0.8–7.5) −4.3 (−8.8 to 0.2) — 
Time >300 mg/dL, % 1.6 (0.0–4.9) 0.6 (0.0–5.5) 0.2 (0.0–2.0) −1.1 (−2.2 to 0.1) — 
Time <70 mg/dL, % 0.00 (0.00–0.25) 0.19 (0.00–0.35) 0.09 (0.05–0.25) 0.03 (−0.01 to 0.08) — 
Time <54 mg/dL, % 0.00 (0.00–0.11) 0.00 (0.00–0.12) 0.02 (0.01–0.08) 0.00 (−0.03 to 0.02) — 
Coefficient of variation, % 29 (27–31) 30 (23–31) 27 (24–30) −1.6 (−4.0 to 0.7) — 

Data are median (IQR) or mean ± SD unless otherwise noted.

FIGURE 1

Cumulative distribution (A), scatter (B), and 24-hour envelope plot (C) of TIR (70–180 mg/dL). In A, the red cumulative distribution during the Control-IQ period shows better TIR compared with the blue cumulative distribution during the CGM-only baseline period. In B, the points above the line of identity denote participants who had better TIR during the Control-IQ period compared with the CGM-only baseline period. Orange dots represent MDI/insulin pump participants, and green dots represent participants using basal insulin only. In C, the line with red symbols denotes hourly median values during the Control-IQ period, and the line with blue symbols denotes hourly median values during the CGM-only baseline period. The shaded regions are defined by the 25th and 75th percentiles.

FIGURE 1

Cumulative distribution (A), scatter (B), and 24-hour envelope plot (C) of TIR (70–180 mg/dL). In A, the red cumulative distribution during the Control-IQ period shows better TIR compared with the blue cumulative distribution during the CGM-only baseline period. In B, the points above the line of identity denote participants who had better TIR during the Control-IQ period compared with the CGM-only baseline period. Orange dots represent MDI/insulin pump participants, and green dots represent participants using basal insulin only. In C, the line with red symbols denotes hourly median values during the Control-IQ period, and the line with blue symbols denotes hourly median values during the CGM-only baseline period. The shaded regions are defined by the 25th and 75th percentiles.

Close modal
FIGURE 2

Mean percentage of TIR (70–180 mg/dL) weekly and by day. The larger graph shows the mean TIR during the CGM-only and open-loop pump baseline and during each week over the 6 weeks of the Control-IQ period. The inset graph shows the mean percentage of TIR during the open-loop pump baseline and during each day for the first 7 days of the Control-IQ period. The solid circles denote the mean values, and the vertical lines extend to ± 1 SE of the mean.

FIGURE 2

Mean percentage of TIR (70–180 mg/dL) weekly and by day. The larger graph shows the mean TIR during the CGM-only and open-loop pump baseline and during each week over the 6 weeks of the Control-IQ period. The inset graph shows the mean percentage of TIR during the open-loop pump baseline and during each day for the first 7 days of the Control-IQ period. The solid circles denote the mean values, and the vertical lines extend to ± 1 SE of the mean.

Close modal

Mean time >180 mg/dL decreased by 15% (95% CI −24 to −5%, P = 0.007), median time >250 mg/dL decreased by 3.7% (95% CI −7.0 to −0.3%, P = 0.03), and mean glucose decreased by 22 mg/dL (95% CI −37 to −6 mg/dL, P = 0.01). Median time <54 mg/dL was 0.00% at baseline and 0.02% during Control-IQ use (mean change = 0.00%, 95% CI = −0.02 to 0.02%).

Both prior MDI regimen/insulin pump users and prior basal-only insulin users showed improvement in CGM metrics (Table 2 and Supplementary Figure S3). Mean TIR increased by 17% in MDI/pump users and by 13% in basal-only insulin users. Results were similar in the 22 participants using an SGLT2 inhibitor or GLP-1 receptor agonist in addition to insulin, for whom TIR increased by 16% and median percentage of time <54 mg/dL changed by 0.00% with Control-IQ (Supplementary Table S6).

There were 252 at-home exercise sessions recorded by 28 of the 30 participants. During exercise and for 2 hours after exercise sessions, mean TIR was 76 ± 32% and median time <70 mg/dL was 0.0% (IQR 0.0–0.0%). From the end of exercise until 6:00 a.m. the next morning, mean TIR was 73 ± 24% and median time <70 mg/dL was 0.0% (IQR 0.0–0.0%).

Mean insulin TDD was 0.62 ± 0.40 units/kg/day at baseline and 0.67 ± 0.36 units/kg/day during Control-IQ use. In prior MDI/pump users, insulin TDD was 0.76 ± 0.41 units/kg/day at baseline and 0.81 ± 0.40 units/kg/day using Control-IQ, whereas in prior basal-only insulin users, insulin TDD was 0.44 ± 0.31 and 0.47 ± 0.17 units/kg/day, respectively. Among the average of 342 boluses delivered per participant by previous MDI users, 52% were autoboluses with the remainder manual boluses, whereas among the average of 327 boluses delivered per participant by prior basal-only insulin users, 53% were autoboluses.

Body weight increased from a median of 81.9 kg (IQR 72.2–107.2 kg) at baseline to 83.2 kg (IQR 74.4–106.5 kg) at 6 weeks (P = 0.003), with the increase observed only in the basal-only insulin users (from a median 79.7 kg at baseline to 81.2 kg at 6 weeks) and not in the MDI/pump users (median 88.5 kg at baseline and 87.1 kg at 6 weeks).

On the PRO surveys, the DIDS satisfaction score increased from 6.9 ± 1.6 to 8.6 ± 1.7 from screening to 6 weeks (P = 0.006), indicating a significant improvement in satisfaction from prior therapy. There were no changes on the DIDS diabetes impact score or the DIDP score. The PROMIS sleep t-score changed from 55.6 ± 7.3 at screening to 52.7 ± 6.4 at 6 weeks (P = 0.19) (Supplementary Table S7). Participants rated the system highly on the System Usability Scale (81.8 of 100), indicating “excellent” usability (10).

There were no episodes of severe hypoglycemia, DKA, HHS, or other serious adverse events.

In this study of adults with type 2 diabetes using basal-bolus or basal-only insulin therapy, use of Control-IQ was safe and associated with a substantial, statistically significant decrease in hyperglycemia resulting in an increase in TIR of 3.6 hours/day (25.2 hours/week). PRO survey results indicated a high degree of satisfaction with Control-IQ compared with prior therapy, an important finding because patient-reported satisfaction is a leading indicator of adherence to treatment interventions (11). The amount of CGM-measured hypoglycemia was low at baseline and remained low during use of Control-IQ. No serious adverse events occurred, and no safety concerns were observed. Although some participants, particularly those not using prandial insulin at the time of enrollment, had weight gain with Control-IQ use, such a finding is not uncommon with glycemic management intensification (12), especially when A1C levels are high (mean A1C was 8.6% in our study cohort), which is indicative of insufficient insulin TDD at baseline. Formal guidance regarding food choices focusing on weight management was not part of this protocol.

The mean TIR during Control-IQ use of 71% is similar to what has been reported in adults with type 1 diabetes (4). Additionally, the median percentage of time participants used the pump with closed-loop functionality was 96%, which is at least as high as what has been observed with type 1 diabetes (4).

The findings in this study are also similar to what has been reported for real-world off-label use of Control-IQ for individuals with type 2 diabetes from the Tandem t:connect Web application. In one analysis, TIR was 76% after 180 days using Control-IQ in 134 prior pump users and 74% in 173 prior MDI users (13). In a second analysis of Control-IQ use for up to 1 year, 378 individuals with type 2 diabetes who switched from Basal-IQ (predictive low-glucose suspension functionality in a t:slim X2 pump) to Control-IQ showed an improvement in TIR from 69% at baseline to 78% after 1 year (14). In an analysis of Medicare and Medicaid recipients who were Control-IQ users, 500 individuals with type 2 diabetes improved from a TIR of 64% at baseline to 72% after using Control-IQ for at least 30 days (15).

Few clinical trials have evaluated an AID system in at-home studies involving people with type 2 diabetes. The fully closed-loop CamAPS HX system has been evaluated in two two-period crossover trials. In 26 adults with type 2 diabetes, Daly et al. (16) reported an increase in TIR from 32 to 66% using the CamAPS HX system for 8 weeks compared with standard insulin therapy without real-time CGM. In 26 adults with type 2 diabetes who were undergoing dialysis, Boughton et al. (17) reported an increase in TIR from 43 to 57% using the CamAPS HX system with fast- acting insulin aspart for 20 days compared with standard insulin therapy without real-time CGM.

Slightly more than half of boluses for prior users of either an MDI regimen or basal-only insulin were given as autoboluses, a unique feature of Control-IQ technology that allows for up to a 6-unit automatic bolus to be given by the system up to once per hour. These autoboluses occur in addition to automated changes in the AID basal rate. In a meta-analysis of randomized trials of Control-IQ users with type 1 diabetes who were 2–72 years of age, Beck et al. (18) showed that the percentage of daily boluses that were autoboluses was 26% for participants with an A1C <7.0% compared with 52% for those with an A1C ≥8.5%. The latter percentage is similar to what was observed in this trial of insulin users with type 2 diabetes. As in the type 1 diabetes analysis, in the current study, the higher the baseline A1C was, the greater the TIR improvement using Control-IQ was. This suggests that autobolusing as a feature of the Control-IQ algorithm has considerable impact in improving A1C levels in individuals with a high A1C before initiating Control-IQ. We speculate that, before the study, missed boluses with an MDI therapy regimen or failure to advance to adding bolus therapy when needed for basal-only insulin users, may explain the elevated A1C and low TIR levels at baseline. Even without changing behavior to initiate the majority of their boluses on their own, participants in the study were able to receive bolus insulin and significantly improve their glycemic outcomes.

The trial was intended to provide a preliminary assessment of Control-IQ use in type 2 diabetes and not definitive results. The short duration and single-arm design of this trial prevented us from assessing whether use of the closed-loop system would lead to persistent weight gain or its longer-term impacts. The sample size was too small to perform any analyses according to C-peptide level. The two participants withdrawn by their site during the pump run-in phase both exhibited a lack of responsiveness to required study contacts and failure to adhere to their diabetes-related medications. It was beyond the scope of this study to optimize the baseline treatment plans of enrolled participants, which may have been affected by medication costs, patient preferences, side effects, and insurance coverage.

The results of this trial demonstrate that, during 6 weeks of use, Control-IQ was safe, with no serious adverse events and no increase in CGM-measured hypoglycemia, with the majority of participants using concurrent GLP-1 receptor agonist and/or SGLT2 inhibitor therapy. There was a substantial improvement in TIR and mean glucose related to a reduction in hyperglycemia. Participant satisfaction improved with Control-IQ use, and the system was rated highly by most participants on the System Usability Scale. The results of this preliminary study support proceeding to conduct a pivotal trial of Control-IQ in adults with type 2 diabetes treated with insulin.

Funding

This study was funded by Tandem Diabetes Care.

Duality of Interest

C.J.L.’s institution has received research support from Abbott Diabetes Care, Dexcom, Insulet, and Tandem Diabetes Care, and she has been a paid consultant to Dexcom. D.R.’s institution has received funding and study supplies from Dexcom and Tandem Diabetes Care. K.P. has received research support from Altimmune, Dexcom, Eli Lilly, Medtronic, Novo Nordisk, Oramed, and Tandem Diabetes Care. T.B. has received research support from Abbott Diabetes Care, Dexcom, Eli Lilly, MannKind, Medtronic, Novo Nordisk, Tandem Diabetes Care, and Viacyte and speaker fees from AstraZeneca, Boehringer Ingelheim, Dexcom, Eli Lilly, and Novo Nordisk. C.M.L.’s institution has received research support from Abbott Diabetes Care, Dexcom, Insulet, and Tandem Diabetes Care. G.O. has received research support, including salary and product support, from Abbott Diabetes Care, Dexcom, Omnipod, and Tandem. J.L.’s, C.K.’s, and R.W.B.’s institutions have received funding and study supplies from Dexcom and Tandem Diabetes Care. No other potential conflicts of interest relevant to this article were reported.

Author Contributions

C.J.L. researched and interpreted the data and wrote the manuscript. D.R. performed statistical analysis and contributed to writing and reviewing the manuscript. Y.C.K., K.P., T.B., L.C., D.D., C.M.L., G.O., C.R., J.L., C.K., and R.W.B. researched and interpreted the data, contributed to discussion, and reviewed/edited the manuscript. C.J.L. is the guarantor of this work and, as such, had full access to all the data in the study and takes responsibility for the integrity of the data and the accuracy of the data analysis.

This article contains supplementary material online at https://doi.org/10.2337/figshare.23925579.

1.
Bergenstal
RM
,
Garg
S
,
Weinzimer
SA
, et al
.
Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes
.
JAMA
2016
;
316
:
1407
1408
2.
Breton
MD
,
Kanapka
LG
,
Beck
RW
, et al.;
iDCL Trial Research Group
.
A randomized trial of closed-loop control in children with type 1 diabetes
.
N Engl J Med
2020
;
383
:
836
845
3.
Brown
SA
,
Forlenza
GP
,
Bode
BW
, et al.;
Omnipod 5 Research Group
.
Multicenter trial of a tubeless, on-body automated insulin delivery system with customizable glycemic targets in pediatric and adult participants with type 1 diabetes
.
Diabetes Care
2021
;
44
:
1630
1640
4.
Brown
SA
,
Kovatchev
BP
,
Raghinaru
D
, et al.;
iDCL Trial Research Group
.
Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes
.
N Engl J Med
2019
;
381
:
1707
1717
5.
Forlenza
GP
,
Lal
RA
.
Current status and emerging options for automated insulin delivery systems
.
Diabetes Technol Ther
2022
;
24
:
362
371
6.
Forlenza
GP
,
Pinhas-Hamiel
O
,
Liljenquist
DR
, et al
.
Safety evaluation of the MiniMed 670G system in children 7–13 years of age with type 1 diabetes
.
Diabetes Technol Ther
2019
;
21
:
11
19
7.
Garg
SK
,
Weinzimer
SA
,
Tamborlane
WV
, et al
.
Glucose outcomes with the in-home use of a hybrid closed-loop insulin delivery system in adolescents and adults with type 1 diabetes
.
Diabetes Technol Ther
2017
;
19
:
155
163
8.
Diabetes Control and Complications Trial Research Group
;
Nathan
DM
,
Genuth
S
,
Lachin
J
, et al
.
The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus
.
N Engl J Med
1993
;
329
:
977
986
9.
Benjamini
Y
,
Hochberg
Y
.
On the adaptive control of the false discovery rate in multiple testing with independent statistics
.
J Educ Behav Stat
2000
;
25
:
60
83
10.
Bangor
A
,
Kortum
P
,
Miller
J
.
Determining what individual SUS scores mean: adding an adjective rating scale
.
J Usability Stud
2009
;
4
:
114
123
11.
Barbosa
CD
,
Balp
MM
,
Kulich
K
,
Germain
N
,
Rofail
D
.
A literature review to explore the link between treatment satisfaction and adherence, compliance, and persistence
.
Patient Prefer Adherence
2012
;
6
:
39
48
12.
Russell-Jones
D
,
Khan
R
.
Insulin-associated weight gain in diabetes: causes, effects and coping strategies
.
Diabetes Obes Metab
2007
;
9
:
799
812
13.
Morberg
JW
,
Singh
H
,
McElwee-Malloy
M
,
Habif
S
,
Constantin
A
.
Real-world evaluation of glycemic outcomes by prior therapy for people with type 1 and type 2 diabetes onboarding to Control-IQ technology [Abstract]
.
Diabetes
2021
;
70
:
710-P
14.
Breton
MD
,
Kovatchev
BP
.
One year real-world use of the Control-IQ advanced hybrid closed-loop technology
.
Diabetes Technol Ther
2021
;
23
:
601
608
15.
Forlenza
GP
,
Carlson
AL
,
Galindo
RJ
, et al
.
Real-world evidence supporting Tandem Control-IQ hybrid closed-loop success in the Medicare and Medicaid type 1 and type 2 diabetes populations
.
Diabetes Technol Ther
2022
;
24
:
814
823
16.
Daly
AB
,
Boughton
CK
,
Nwokolo
M
, et al
.
Fully automated closed-loop insulin delivery in adults with type 2 diabetes: an open-label, single-center, randomized crossover trial
.
Nat Med
2023
;
29
:
203
208
17.
Boughton
CK
,
Tripyla
A
,
Hartnell
S
, et al
.
Fully automated closed-loop glucose control compared with standard insulin therapy in adults with type 2 diabetes requiring dialysis: an open-label, randomized crossover trial
.
Nat Med
2021
;
27
:
1471
1476
18.
Beck
RW
,
Kanapka
LG
,
Breton
MD
, et al
.
A meta-analysis of randomized trial outcomes for the t:slim X2 insulin pump with Control-IQ technology in youth and adults from age 2 to 72
.
Diabetes Technol Ther
2023
;
25
:
329
342
Readers may use this article as long as the work is properly cited, the use is educational and not for profit, and the work is not altered. More information is available at https://www.diabetesjournals.org/journals/pages/license.