Background: The Eversense CGM System sensor is implanted in the upper arm. It uses a fluorescent, boronic acid glucose-binding moiety to measure glucose. The glucose indicator is susceptible to oxidative de-boronation from reactive oxygen species (ROS) in the interstitial fluid, necessitating daily calibrations. To reduce calibration frequency, a ROS indicator was added to measure oxidative de-boronation. Figure 1 (a) shows the copolymerization of glucose and ROS indicators into a single hydrogel matrix with orthogonal excitation emission properties. The sensor uses a 380 nm LED for excitation of the glucose indicator and a 470 nm LED for excitation of the ROS indicator.

Methods: The fluorescence of the indicators was characterized in vitro. In addition, sensors were implanted in subjects with type 1 diabetes in a 365-day clinical feasibility study to assess accuracy against fingerstick measurements.

Results: Figure 1 (b) illustrates the relationship whereby ROS exposure increases the fluorescence of the ROS indicator and decreases the fluorescence of the glucose indicator. A MARD of 9.9% against fingerstick with one calibration per week was observed through 90 days.

Conclusions: Early data suggests the multiple-analyte system should allow the next generation of long-term, implanted CGM systems to maintain accuracy with significantly reduced calibration requirements.

Disclosure

V.N.Velvadapu: None. J.M.Masciotti: Employee; Senseonics. A.D.Dehennis: Employee; Senseonics. P.Sanchez: Employee; Senseonics. K.S.Tweden: Employee; Senseonics. F.R.Kaufman: Consultant; Twin Health, Employee; Senseonics. S.Ioacara: Research Support; Senseonics.

Readers may use this article as long as the work is properly cited, the use is educational and not for profit, and the work is not altered. More information is available at http://www.diabetesjournals.org/content/license.