Background: Multiple pivotal and post market studies have confirmed the safety and accuracy of implantable Eversense CGM Systems. A post market study was initiated to compare Eversense 90-day and 180-day systems to SMBG in CGM naïve patients. Methods: This post market study is a prospective, multicenter, one year comparison of SMBG to Eversense in CGM naïve adults with diabetes in US sites (up to 925 users). After baseline assessment including HbA1c, patients used SMBG for 6 months (phase 1) followed by Eversense for 6 months (phase 2). Sensors were inserted at the start of phase 2. Visits occurred q 90 days to collect SMBG or CGM data and assess adverse events. HbA1c was measured at 6 and 12 months. Glucometrics were calculated in both phases. Enrollment and follow-up visits are ongoing.

Results: In 15 sites, 84 users have completed the study (mean 47.6% male, 90.5% type 2 diabetes, 57.2 years age). In the initial users, reductions in HbA1c were seen in both phases and improvements in glucometrics were observed (Table) with CGM compared to SMBG.

Conclusion: In a CGM naïve population, 6 months of SMBG use reduced HbA1c. There was a further significant reduction after 6 months of using Eversense CGM, as well as significant improvement in glucometrics - an increase in Time in Range (70-180 mg/dL) with no increase Time below Range. These data suggest that Eversense CGM improves glucose management compared to SMBG.

Disclosure

K.S.Tweden: Employee; Senseonics. B.Romarowski: Employee; Senseonics. C.Mdingi: Employee; Senseonics. F.R.Kaufman: Consultant; MannKind Corporation, Twin Health, Stock/Shareholder; Senseonics.

Readers may use this article as long as the work is properly cited, the use is educational and not for profit, and the work is not altered. More information is available at http://www.diabetesjournals.org/content/license.