Full realization of the benefits of continuous glucose monitoring (CGM) depends on addressing barriers such as cost, accuracy, burdens of daily use, and uncertainty about applying the data. Lack of systematic education has also hampered widespread adoption among patients and health care professionals. This article describes the practical application of an affordable and intuitive category of CGM called “flash” that requires users to scan the sensor with a handheld reader for on-demand access to continuous data. The data may be used for in-the-moment therapy adjustment, retrospective review of glucose patterns, and observation of glucose trending in response to behavior. Higher rates of flash CGM scanning have been associated with increased time in the glycemic target range and reduced time in hyper- and hypoglycemia. Growing interest in this technology suggests new opportunities for helping more patients incorporate CGM into their daily self-care.

The health care improvement goal of excellent quality and patient experience at reasonable cost has translated to new expectations of glucose measurement. A1C is the reigning standard for assessing long-term diabetes complications risk, but as a measure of average glucose level over 2–3 months, this laboratory assay cannot capture daily glucose change and is subject to inaccuracies in individual patients (1). Self-monitoring of blood glucose (SMBG) is typically used to supplement A1C because episodic fingerstick checking allows patients to assess their response to factors immediately affecting daily glucose change. However, interrelated issues of education, logistics, and reimbursement often impede patient engagement with a structured monitoring schedule (26).

In the past decade, continuous glucose monitoring (CGM) has emerged as an adjunct to A1C that could potentially help meet glycemic goals (710). CGM technology provides near-continuous data by measuring the glucose concentration in the body’s interstitial fluid and extrapolating blood glucose levels for real-time or retrospective analysis.

Clinical trials have affirmed the benefit of accessing continuous data on a regular basis. Specifically, study participants have experienced improved A1C, decreased time spent in hyper- and hypoglycemia, and lower incidences of severe hypoglycemia (1122). These benefits are derived from the ability of CGM, compared to A1C and SMBG, to more fully communicate real-time glucose values, trend information, and potentially harmful high and low glucose swings known as glycemic variability (2,23). Users wearing traditional devices can receive alerts to actual or predicted episodes of hypo- and hyperglycemia and review display arrows reflecting the rate and direction of glucose change.

Despite these advantages, CGM use remains at only 8–17%, even among motivated patients with type 1 diabetes (2427). Longstanding barriers have included high cost, concerns about accuracy, alarm fatigue, encumbrances of wearing a mechanical device, uncertainty about applying the data, and the continuing need for SMBG to dose insulin (15,2830). Additionally, systems to date have been education-intensive, and few clinicians or patients receive systematic training on how to respond to readings, either in the moment or retrospectively (28).

This article examines evolving CGM use in the wake of newly introduced flash CGM (FCGM) technology. This novel category of continuous data capture can add context to A1C, eliminate the need for routine fingerstick monitoring, and provide “teaching moments” amenable to wider patient engagement with CGM-guided diabetes management.

FCGM is available in two versions, personal and professional. The personal system, FreeStyle Libre (Abbott Diabetes Care, Alameda, Calif.), was approved as a replacement for fingerstick checking by the U.S. Food and Drug Administration (FDA) in September 2017. It affords on-demand observation of real-time and trend data, as well as retrospective review of complete profiles by patients at home or providers in their clinics (31). The earlier approved FreeStyle Libre Pro system (Abbott Diabetes Care, Alameda., Calif.) is referred to as a professional device because patients wear it without being able to see glucose values until their provider uploads the data for review during an office visit. Both versions include a disposable sensor worn on the back of the arm for up to 14 days (32).

Although employing the same chemical glucose oxidase mechanism for glucose measurement as traditional CGM, FCGM uses wired enzyme glucose sensing technology to automatically measure glucose every minute and records readings at 15-minute intervals. It is calibrated at the time of manufacturing and does not require recalibration by patients or clinicians. Additionally, there is no separate transmitter that must be reused with sensors. These features reduce patient burden and enhance ease of use.

Mechanism of Action

Personal and professional FCGM provide different perspectives for making therapeutic or behavioral adjustments that affect glycemic control (3335). Personal FCGM enables patients to make “micro” adjustments through information obtained by scanning, or “flashing,” the glucose sensor with a handheld reader or phone app. The sensor, about the size of two stacked U.S. quarters, can be swiped through clothing for displays of last-minute glucose values, trend arrows, and graphs showing the last 8 hours of data. There is no limit to the number of scans that can be taken over the duration of a sensor use.

For many patients who do not perform SMBG regularly, personal FCGM may afford the first opportunity to observe glucose trending in response to behavior, such as how glucose rises after eating or falls after exercise (3638). Personal FCGM may also benefit people with insulin-requiring diabetes who rely on continuous data for glucose management but struggle with certain aspects of traditional CGM systems, such as alarms (39). Candidates include people with newly diagnosed diabetes using SMBG with mixed success, as well as those on intensive insulin regimens who do not have impaired awareness of hypoglycemia. Sensor readings can be used for insulin dosing unless glucose is changing rapidly, physical symptoms do not match the values on the reader, or there is a “check blood glucose” alert on the home screen (32). In such situations, confirming sensor values with fingerstick checks is vital because changes in plasma glucose precede changes in interstitial glucose and thus provide an earlier warning of impending hypo- or hyperglycemia.

The professional version of FCGM, unlike the personal version, is designed only for retrospective review of glycemic patterns that inform “macro” adjustments of therapies and behaviors, such as introducing a new medication or changing a patient’s insulin-to-carbohydrate ratio. As previously mentioned, patients have no interaction with the device and cannot see glucose readings during wear (40,41). Data can be retrieved only when the health care professional scans the reader over the sensor and uploads the information to the LibreView desktop software (42). At that point, the clinician or educator, usually together with the patient, can view a single-page report, described later in this article, to quickly identify trouble spots such as postprandial hyperglycemia or explain A1C values that are higher or lower than would be expected based on SMBG readings alone (1).

Patients who have had no experience with CGM may most likely rely on their diabetes care team to recommend a particular system (43). Factors to consider when matching CGM technologies to patients’ individual needs include accuracy, convenience, data display formats, cost, and the learning curve for both patients and clinicians (44). Selecting a continuous monitoring device should be a deliberate decision that balances the operational and design differences among systems with the clinical, treatment, and lifestyle needs of the patient. Device attributes for guiding the selection are presented in Table 1 (4547).

TABLE 1.

Features of Selected CGM Systems (4547)

SystemWear TimeWarm-Up TimeCalibrationFrequency of Glucose ReadingsAccuracy (MARD)Software and/or Device Compatibility
Professional systems 
Freestyle Libre Pro Up to 14 days 1 hour None Every 15 minutes 12.3% Data scanned in the provider’s office from sensor using Freestyle Libre Pro reader in office 
Medtronic iPro 2 Enlite CGM sensor and iPro2 digital recorder Up to 6 days 1 hour None, but at least one blood glucose entry every 12 hours is required for system uploads Every 5 minutes 13.6% Data uploaded from sensor recorder using Medtronic CareLink iPro website 
Dexcom G4 Platinum professional system 7 days 2 hours Every 12 hours after the 2-hour start-up calibration Every 5 minutes 13.3% Data uploaded from sensor using Dexcom STUDIO software 
Personal systems 
FreeStyle Libre Up to 14 days 1 hour None Available every minute; automatically records every 15 minutes 9.4% Data may be uploaded from reader in provider’s office using Freestyle LibreView software 
Dexcom Platinum G4/G5 sensor with G4 Platinum transmitter Up to 7 days 2 hours Every 12 hours Every 5 minutes 9% when used with most current Dexcom software Stand alone with Dexcom G4 receiver and compatible with Animas Vibe and Tandem t:slim insulin pumps 
Dexcom Platinum G4/G5 sensor with G5 Platinum transmitter Up to 7 days 2 hours Every 12 hours Every 5 minutes 9% when used with most current Dexcom software Stand alone with Dexcom G5 receiver, most Apple and Android products, and compatible with Tandem t:slim X2 insulin pump 
Dexcom G6 sensor and transmitter Up to 10 days 2 hours None Every 5 minutes 9.8% overall and 9.6% in children 6–17 years of age Stand alone with most Dexcom G5 receivers and G6 receiver, most Apple and Android products 
Medtronic Enlite sensor and MiniLink or Guardian Link transmitter Up to 6 days 2 hours Every 12 hours Every 5 minutes 13.6% Compatible with Medtronic 530G and 630G insulin pumps 
Medtronic Guardian Sensor 3 sensor and Guardian Link 3 transmitter Up to 7 days 2 hours Every 12 hours Every 5 minutes 10.6% with 2 calibrations/day and 9.6% with 3–4 calibrations/day Compatible with Medtronic 670G hybrid closed-loop insulin pump system 
SystemWear TimeWarm-Up TimeCalibrationFrequency of Glucose ReadingsAccuracy (MARD)Software and/or Device Compatibility
Professional systems 
Freestyle Libre Pro Up to 14 days 1 hour None Every 15 minutes 12.3% Data scanned in the provider’s office from sensor using Freestyle Libre Pro reader in office 
Medtronic iPro 2 Enlite CGM sensor and iPro2 digital recorder Up to 6 days 1 hour None, but at least one blood glucose entry every 12 hours is required for system uploads Every 5 minutes 13.6% Data uploaded from sensor recorder using Medtronic CareLink iPro website 
Dexcom G4 Platinum professional system 7 days 2 hours Every 12 hours after the 2-hour start-up calibration Every 5 minutes 13.3% Data uploaded from sensor using Dexcom STUDIO software 
Personal systems 
FreeStyle Libre Up to 14 days 1 hour None Available every minute; automatically records every 15 minutes 9.4% Data may be uploaded from reader in provider’s office using Freestyle LibreView software 
Dexcom Platinum G4/G5 sensor with G4 Platinum transmitter Up to 7 days 2 hours Every 12 hours Every 5 minutes 9% when used with most current Dexcom software Stand alone with Dexcom G4 receiver and compatible with Animas Vibe and Tandem t:slim insulin pumps 
Dexcom Platinum G4/G5 sensor with G5 Platinum transmitter Up to 7 days 2 hours Every 12 hours Every 5 minutes 9% when used with most current Dexcom software Stand alone with Dexcom G5 receiver, most Apple and Android products, and compatible with Tandem t:slim X2 insulin pump 
Dexcom G6 sensor and transmitter Up to 10 days 2 hours None Every 5 minutes 9.8% overall and 9.6% in children 6–17 years of age Stand alone with most Dexcom G5 receivers and G6 receiver, most Apple and Android products 
Medtronic Enlite sensor and MiniLink or Guardian Link transmitter Up to 6 days 2 hours Every 12 hours Every 5 minutes 13.6% Compatible with Medtronic 530G and 630G insulin pumps 
Medtronic Guardian Sensor 3 sensor and Guardian Link 3 transmitter Up to 7 days 2 hours Every 12 hours Every 5 minutes 10.6% with 2 calibrations/day and 9.6% with 3–4 calibrations/day Compatible with Medtronic 670G hybrid closed-loop insulin pump system 

The chief distinction between FCGM and traditional CGM is that FCGM does not passively send continuous data to receivers, phones, or a pump. Instead, users must actively scan their sensor with a reader to see their glucose information displayed. For some patients, repeated scanning could potentially increase engagement with CGM-driven diabetes management (48). FCGM also differs from traditional CGM in that the U.S. version of the device does not feature programmable alerts and alarms that warn patients of current or impending hyper- or hypoglycemia. This aspect offers an acceptable option for patients interested in acquiring continuous data but who are concerned about false or frequent alarms (44). Despite having no alarms, the FCGM system provides a distinct audible tone when alerts to perform fingerstick checks are displayed. This tone may occur during scanning when glucose is <70 or >240 mg/dL, projected to be <70 or >240 mg/dL, identified as “hi” or “lo” or projected to be “hi” or “lo,” or rapidly changing or when no trend arrow displays. SMBG remains a required action under conditions of rapidly changing glucose, when clinical signs are inconsistent with displayed values, for confirmation of sensor-reported hypoglycemia, and during the 1-hour warm-up period while the device is adjusting to the body (49).

The retail price of FCGM is currently less than that of traditional systems (29). As with the Dexcom G5 system (Dexcom, San Diego, Calif.), FCGM is covered under Medicare for people using insulin who meet eligibility criteria (see below) (50). Finally, FCGM is not affected by acetaminophen interference in contrast with many newer-generation CGM devices (51).

Patients’ continued use of CGM for diabetes management is directly related to their trust in the accuracy and reliability of data (5254). Whereas ISO (International Organization for Standardization) 15197 is the accepted accuracy standard for SMBG devices, mean absolute relative difference (MARD) is the most common metric used to assess the performance of CGM systems. This metric shows on average how far away the sensor reading is from a reference blood glucose value across all glucose ranges. MARD values are expressed as a percentage; the lower the percentage value, the more accurate the system.

A 2015 performance study using the aggregate MARD between all sensor values and matched reference values reported good accuracy of FCGM against capillary and venous glucose measurements in a wide range of patients with type 1 or type 2 diabetes (55). Although comparing MARD values among CGM systems is difficult because of a lack of standardization among clinical study methodologies and differences between older and newer-generation devices, FDA assessments for product approval indicate comparable accuracy among currently available systems (30,5658).

CGM Comparison Studies

Head-to-head accuracy comparisons between FCGM and selected newer-generation traditional systems have shown similar accuracy. A recent study by Aberer et al. (59) comparing the FreeStyle Libre with the Dexcom G4 Platinum and Medtronic MiniMed 640G (Medtronic Diabetes, Northridge, Calif.) systems over 12 hours (24 hours after sensor insertion) during mimicked real-life conditions such as meals, exercise, and hypo- and hyperglycemia found that MARDs overall were 13.2% (± 10.9%), 16.8% (± 12.3%), and 21.4% (± 17.6%), respectively (59). All three sensors performed less accurately during hypoglycemia and best during hyperglycemia, although comparative performance among systems can vary significantly and inexplicably among patients (60). It is therefore important to confirm sensor readings with blood glucose measurements in situations where glucose is rapidly changing or in the hypoglycemic range (<70 mg/dL) or when symptoms are inconsistent with the value on the display screen (49).

To document the performance of the FreeStyle Libre and Dexcom G4 systems during glycemic excursions, Boscari et al. (61) collected accuracy data from 22 adults with type 1 diabetes both at home and during a single 6-hour hospital admission to induce glycemic excursions (early post-meal hyperglycemia followed by a quick decrease in blood glucose). The sensors functioned with comparable accuracy during home use, but the accuracy for both systems decreased in the hypoglycemic range owing to the lag time between plasma and interstitial glucose (62). A follow-up study (63) with the newer-generation Dexcom G5 Mobile sensor (Dexcom, San Diego, Calif.), which like the FreeStyle Libre is approved for making diabetes treatment decisions without the need for confirmatory fingerstick checking, found that both systems performed safely and effectively, with an overall at-home MARD of 12.3% for the FreeStyle Libre and 9.8% for the G5 (P <0.001) (63). However, the MARD increased during hypoglycemia and decreased during hyperglycemia with both systems, again pointing to the need for confirming CGM with SMBG when results are in the hypoglycemic range or inconsistent with symptoms.

A consistent finding of all trials of personal CGM is that frequent checking or scanning of the device is essential for successful outcomes (30). Dunn et al. (48) evaluated de-identified data for 50,831 FreeStyle Libre system readers with 279,446 sensors (86.4 million hours of readings and 63.8 million scans). Scan rate per reader was determined and readers were sorted into 20 equal-sized, rank-ordered user groups distinguished by scan frequency. Estimated A1C and other glucose parameters were calculated for each group. Analysis revealed an average scan rate of 16.3 times/day. Estimated A1C levels decreased (P = 0.001) from 8.0% in the group with the lowest scan rate (4.4 times/day) to 6.7% for those with the highest scan rate (48 times/day). Scan rates also correlated with hypoglycemia rates; as scan rates increased, hypoglycemia rates <70, <55, and <45 mg/dL decreased by 15, 40, and 49%, respectively (all P <0.001). Furthermore, time in range, defined as glucose levels between 70 and 180 mg/dL, increased from 12 to nearly 17 hours/day comparing the lowest with the highest scan rate users, representing a 40% increase in time spent in euglycemia. The authors speculated that the ease of checking glucose may have contributed to the high frequency of scanning.

The effect of FCGM on hypoglycemia reported by Dunn et al. (48) is consistent with an earlier study by Bolinder et al. (64), which compared FCGM to SMBG in European adults with well-controlled type 1 diabetes (n = 239) (64). Participants in the FCGM group spent 38% less time in the hypoglycemic range (<70 mg/dL). This reduction was accomplished with no change in total daily insulin dose or deterioration of A1C. Time in range significantly increased in the intervention group; high scores for treatment satisfaction and a scan rate averaging 15 scans/day indicated a high level of acceptance of FCGM consistent with subsequent reports of patient experience (65).

Two single-arm studies without control groups demonstrated significant A1C improvement after FCGM initiation. In the first study, Dover et al. (66) prospectively evaluated FCGM in 25 participants with type 1 diabetes and reported improved glucose control, fewer episodes of hypoglycemia, and improved quality of life. Mean A1C fell from 8.0 ± 0.14% to 7.5 ± 0.14% (–0.48%, P = 0.001) after 16 weeks of FCGM. The number of people with an A1C of ≤7.5% more than doubled after FCGM use. Those with a baseline A1C >7.5 experienced a greater reduction than participants with an A1C <7.5% at baseline (–0.59 ± 0.15% vs. –0.2 ± 0.11%, P = 0.005). FCGM data showed that the number of hypoglycemic episodes (<72 mg/dL) dropped from 17 in the first 2 weeks of use to 12 in the final 2 weeks. The second study by Ish-Shalom et al. (67) reported similar outcomes in patients with difficult-to-control type 1 (n = 6) or type 2 (n = 25) diabetes. A1C decreased by 1.33 ± 0.29% after 8 weeks, and for those who continued using FCGM after the 12-week study period (n = 27), the change was sustained for 24 weeks (1.21 ± 0.42%, P = 0.009). Questionnaires completed by all 31 participants indicated high satisfaction and desire to continue using the device.

In a large multicenter study of patients with type 2 diabetes (n = 224) on intensive insulin therapy, Haak et al. (68) compared FCGM to standard fingerstick checking. A1C reductions in the FCGM and SMBG groups were comparable overall (–0.29 ± 0.07% in the intervention group vs. –0.31 ± 0.09% in the control group); however, patients <65 years of age in the FCGM group showed significant A1C improvement compared to the control group (–0.53 ± 0.09% vs. –0.20 ± 0.12%, P = 0.0301). Times in hypoglycemia <70 mg/dL and <55 mg/dL were reduced 0.47 ± 0.13 hours/day (43%) and 0.22 ± 0.07 hours/day (53%), respectively, for FCGM versus SMBG users. Nocturnal hypoglycemia (<70 mg/dL) declined by 54% in the FCGM group (P = 0.0001), and time in hypoglycemia was reduced by 56% (P = 0.0083) for patients ≥65 years of age. Treatment satisfaction was higher in FCGM users, and no device-related serious adverse events were reported.

To meet the triple aim of providing high-quality patient experience, wider population health, and reduced cost of care, it will be necessary to break the gridlock of suboptimal glycemic control, defined as an A1C >7.0%, for almost 50% of patients diagnosed with diabetes (8,9,69). In part, this will come about by optimizing continuous data to “unmask” specific glucose dynamics behind A1C, which remains the gold standard for estimating the risk of long-term complications and mortality but does not reveal intra- and inter-day variations that may lead to hypoglycemia or postprandial hyperglycemia. Furthermore, the range of average glucose corresponding to an A1C level increases at each successive increment such that the average glucose of an individual with an A1C of 7.0% (95% CI 123–185 mg/dL) could in fact be higher than the average glucose of another with an A1C of 8% (95% CI 147–217 mg/dL) (70). The presence of hemolytic anemia, hemoglobinopathy, and inter-individual variations in red blood cell life span also may falsely raise or lower results. If unrecognized by patients and their health care providers, these limitations of A1C interpretation have the potential to set up a cycle of discouragement and disengagement when targets remain unmet despite patients’ best efforts at daily self-management (71).

Role of Continuous Glucose Data

Easier, more intuitive monitoring technologies such as FCGM portend a shift toward wider use of continuous glucose data to better decipher A1C and guide daily adjustments in therapy (2,23,28,30,72). The more often glucose is within the target range, the better the A1C is likely to be because these two variables are highly correlated (73). In a 2015 white paper, the American Association of Diabetes Educators advocated CGM for any person with diabetes who is willing to wear a device, regardless of diabetes type, duration, or the age of the patient (43). Main purposes cited by the expert panel were identifying glycemic excursions, validating therapy adjustments, observing and modulating the effects of physical activity and meals on glucose levels, and using trend information to prevent or mitigate issues associated with fluctuation of blood glucose known as glycemic variability, a strong predictor of hypoglycemia and poor glycemic management (19,7484).

Glucose metrics that play a role in meeting these goals—including time in range, time in hypoglycemia, time in hyperglycemia, and glucose variability—are best calculated by CGM (30,71). Because FCGM poses comparatively fewer demands than traditional CGM, it offers an opportunity to head off historical obstacles when introducing continuous data, especially to patients who have performed SMBG erratically, unsuccessfully, or not at all (48,66). The rest of this article will describe mechanical and data features of FCGM to help guide patients in using this technology for daily diabetes self-management.

The FreeStyle Libre system consists of a handheld reader and fully disposable sensor designed for upper arm wear (85). The sensor can be applied to the upper arm with one hand by pushing the applicator down firmly on the prepared site (swabbed with alcohol and air dried). The patient initiates the sensor with the reader and can begin obtaining glucose readings after 1 hour. The sensor may be worn for up to 14 days.

Before starting personal FCGM, patients are advised to review all of the product instructions and other educational materials provided with the product and on the manufacturer’s website. Scanning frequently can help patients see how carbohydrate intake, medication, illness, exercise, or stress affect readings. Optimal scanning times can be established, and the clinician or educator can address the issue of over-responding to above-target post-meal CGM readings. Taking rapid-acting insulin at close intervals, commonly referred to as “insulin stacking,” is a well-known cause of hypoglycemia, and correction doses are rarely required within 2 hours of a meal dose.

Direct questioning can help determine whether patients using rapid-acting insulin understand the principles of mealtime and correction dosing. Questions may include: “How much time do you wait to eat after taking your mealtime insulin?”, “How long do you wait between insulin doses to avoid insulin stacking?”, “How often do you check your glucose before bed?”, and “When should you eat a bedtime snack?” Showing patients the Notes feature (a pencil symbol in the upper right corner of the reader home screen) can encourage systematic tracking of insulin, food, and other factors affecting glucose levels. Food and rapid-acting insulin notes, if saved, will be available for review by the clinician or educator.

There are several situations in which sensor readings alone should not be used for making treatment decisions. These include rapidly changing glucose levels indicated by trends arrows pointing straight up or down, sensor readings in the hypoglycemic range, or when symptoms do not match sensor readings. In these instances, confirmatory blood glucose checking should be performed.

Potential skin reactions should also be discussed upon initiation of personal or professional FCGM. In one study by the manufacturer, mild skin irritations were reported in 5 of 48 patients (85). When the sensor is applied, a small filament is inserted, usually painlessly, just under the skin and held in place with an adhesive pad. To avoid irritation, patients may select a different site on the back of the arm when replacing the sensor. There is no need to stretch or pinch the skin. The waterproof sensor can be worn while bathing, showering, swimming in depths <3 feet for under 30 minutes, or exercising.

Considerations for Professional FCGM

Because the FCGM sensor can be worn for up to 14 days without replacement and there is no need for calibration or patient interaction with the device, a professional study affords a detailed view of a patient’s changing glucose levels. The use of professional FCGM will vary from practice to practice depending on size, logistics, and resources. Training one member of the team to specialize in professional FCGM start-ups, application, downloading, and report generation can facilitate workflow. Nonmedical staff can confirm a patient’s insurance status and schedule follow-up. Payer policy and criteria should be confirmed in the event that multiple professional FCGM studies are required to follow a patient’s progress over the course of a given year.

Start-ups and downloads of professional FCGM are usually accomplished in two visits. At the first, the health care professional counsels the patient about the reasons and procedures for using professional FCGM and inserts the sensor. After the 14-day wear period, the patient returns for sensor removal, report downloading, and data interpretation by a qualified health care professional (see Using FCGM With Data Management Tools below).

Reimbursement

As evidence of increasing acceptance of CGM, a Current Procedural Terminology code (95249) was recently designated for procedures surrounding personal use devices worn for at least 72 hours, including sensor placement, calibration (if applicable), patient training, and a printout of the recording. In the case of professional CGM, coverage usually includes sensor application and patient counseling at the first visit and removal of the sensor and report downloading at the second visit (code 95250). Report interpretation is covered separately (code 95251).

As mentioned previously, Medicare patients with type 1 or type 2 diabetes may be eligible for coverage of nonadjunctive personal CGM if they currently perform at least four fingersticks per day with a home blood glucose meter; they use insulin, either with multiple daily injections or an insulin pump; and they will need to make frequent adjustments in insulin doses based on CGM readings (50).

Patients using the personal version of FCGM carry the reader to scan the sensor for nearly immediate feedback. The sensor must be scanned at least three times a day, 8 hours apart, for complete data capture. From the home screen, the user may add tags to each scan (e.g., carbohydrates consumed, insulin taken, or exercise performed) or access a glucose history from the past 90 days (see below). It should be noted that when >8 hours pass between scans, the oldest data are overwritten by the most recent data. For this reason, people who wake during the night are encouraged to scan to maintain a continuous report. Additionally, because FCGM does not feature automatic alarms for nocturnal hypoglycemia, checking current sensor glucose and trend arrow data before sleep is recommended.

Patients accessing FCGM data in real time should be urged to use all of the information on the display screen when making treatment decisions. For many people, however, the trend-arrow feature of FCGM is often the most illuminating (86). The directionality of trend arrows enables users to anticipate future glucose concentrations. This foresight can be used to proactively adjust therapy and prevent hypo- or hyperglycemia (87).

Patients who have no experience with continuous data might elect to incorporate trend arrow–guided decision-making gradually as they gain improved understanding of how circumstances such as meals, physical activity, and insulin on board affect their individual glucose response. Arrows for personal FCGM are defined as:

  • ↑ rising quickly (>2 mg/dL/minute);

  • ↗ rising (between 1 and 2 mg/dL/minute);

  • → changing slowly (<1 mg/dL/minute);

  • ↘ falling (between 1 and 2 mg/dL/minute); and

  • ↓ falling quickly (>2 mg/dL/minute).

An example of how trend arrows can help in making treatment decisions is shown in Table 2.

TABLE 2.

Sample of Trend Arrow–Guided Decision-Making (87)

Patient ProfileScanning TimeWhat Display ShowsWhat the Patient Does
Kate has a target of 100 mg/dL and a correction factor of 1:50. This means she would take 1 unit of insulin to lower her glucose about 50 mg/dL. After breakfast  Kate sees a reading of 250 mg/dL trending rapidly downward. There is also a high glucose message and the “check blood glucose” symbol.  Seeing the symbol, Kate performs a blood glucose check before deciding what to do. 
Before lunch  Kate’s glucose is 250 mg/dL and rising. Before eating, Kate adds 50 mg/dL to her current reading given the rising trend arrow (250 + 50 = 300). She subtracts her target (300 –100 = 200) and divides by her correction factor (200 ÷ 50 = 4). Kate takes 4 units of insulin in addition to her calculated meal dose. 
After lunch  Ninety minutes later, Kate’s glucose is the same. The trend arrow and graph show a continued rise. Kate does not take a correction dose because it is within 2 hours of her meal dose. This could lead to “insulin stacking” and low glucose. The insulin she took for her meal may still be active. Kate decides to wait and scan again later. 
Before dinner  Kate’s current glucose is 250 mg/dL. The trend arrow and graph indicate her glucose is going down. Kate asks herself what might be causing her glucose to go down and what she might do to prevent a low glucose, deciding to take less insulin before her meal. She subtracts 50 mg/dL from the current value because of the falling trend arrow (250 – 50 = 200) and then subtracts her target of 100 mg/dL (200 – 100 = 100). She divides this by her correction factor (100 ÷ 50 = 2). Kate takes 2 units of insulin in addition to her calculated meal dose. 
Patient ProfileScanning TimeWhat Display ShowsWhat the Patient Does
Kate has a target of 100 mg/dL and a correction factor of 1:50. This means she would take 1 unit of insulin to lower her glucose about 50 mg/dL. After breakfast  Kate sees a reading of 250 mg/dL trending rapidly downward. There is also a high glucose message and the “check blood glucose” symbol.  Seeing the symbol, Kate performs a blood glucose check before deciding what to do. 
Before lunch  Kate’s glucose is 250 mg/dL and rising. Before eating, Kate adds 50 mg/dL to her current reading given the rising trend arrow (250 + 50 = 300). She subtracts her target (300 –100 = 200) and divides by her correction factor (200 ÷ 50 = 4). Kate takes 4 units of insulin in addition to her calculated meal dose. 
After lunch  Ninety minutes later, Kate’s glucose is the same. The trend arrow and graph show a continued rise. Kate does not take a correction dose because it is within 2 hours of her meal dose. This could lead to “insulin stacking” and low glucose. The insulin she took for her meal may still be active. Kate decides to wait and scan again later. 
Before dinner  Kate’s current glucose is 250 mg/dL. The trend arrow and graph indicate her glucose is going down. Kate asks herself what might be causing her glucose to go down and what she might do to prevent a low glucose, deciding to take less insulin before her meal. She subtracts 50 mg/dL from the current value because of the falling trend arrow (250 – 50 = 200) and then subtracts her target of 100 mg/dL (200 – 100 = 100). She divides this by her correction factor (100 ÷ 50 = 2). Kate takes 2 units of insulin in addition to her calculated meal dose. 

Users seeking additional insight into current glucose values and trends can select the “review history” menu option for 7-, 14-, 30-, and 90-day scroll-through reports of average glucose; time-in-target trends; daily patterns of hypo- and hyperglycemia; low glucose events; and frequency of scanning.

Teaching patients with diabetes the thought process behind continuous glucose data analysis can lead to more strategic and independent glucose management (23,30). To this end, the ambulatory glucose profile (AGP) facilitates interpretation and comprehension of continuous glucose data. FCGM personal and professional devices were among the first to feature the AGP, which is now available with other CGM systems as well (88,89).

AGP software collapses and plots all collected glucose values as if they occurred in a single 24-hour period. The resulting graph (Figure 1) has been incorporated in a standardized report, which has been endorsed by the American Association of Clinical Endocrinologists for continuous glucose data reporting (Figure 2). This standardized report begins with a statistical summary showing glucose exposure, glucose variability (coefficient of variation [CV] and standard deviation [SD]), the proportion of glucose values in target range (70–180 mg/dL), the percentage of values above or below target (low, serious low, high, and serious high), and the percentage of time CGM is active. Beneath this summary, the AGP is displayed with five distribution curves drawn from the aggregated glucose readings to provide an at-a-glance picture of a standard 24-hour period. The dark blue line represents the median curve and would be mostly flat under optimal conditions. The curves immediately above and below the median curve (25th and 75th percentiles) depict the daily, nightly, and postprandial excursions for 50% of the aggregated glucose values; a wider span (indicated by blue shading) indicates higher risk for glycemic variability during the associated time period, whereas a narrower span denotes lower risk. The dashed curves represent the 10th and 90th percentiles, showing data above or below 80% of all the data (indicated by gray shading), conveying “occasional excursions.”

FIGURE 1.

AGP report. The AGP is a visual report that collapses all glucose readings from several days or weeks as if they occurred in a single 24-hour period, making it easier to visualize glycemic patterns. The median (dark blue line) shows the middle of the data; the 25th and 75th percentile curves (blue shading), or interquartile range, represent 50% of all the data; the 10th and 90th percentile curves (gray shading) signify 80% of all data. The 10th and 90th percentiles showing data above or below 80% of all the data (indicated by edges of gray shading) convey “occasional excursions.”

FIGURE 1.

AGP report. The AGP is a visual report that collapses all glucose readings from several days or weeks as if they occurred in a single 24-hour period, making it easier to visualize glycemic patterns. The median (dark blue line) shows the middle of the data; the 25th and 75th percentile curves (blue shading), or interquartile range, represent 50% of all the data; the 10th and 90th percentile curves (gray shading) signify 80% of all data. The 10th and 90th percentiles showing data above or below 80% of all the data (indicated by edges of gray shading) convey “occasional excursions.”

Close modal
FIGURE 2.

AGP report notes for a 24-year-old woman who has had type 1 diabetes for 2 years and is beginning FCGM. Her A1C is 7.8%. A) Statistical summary shows high average glucose and high CV (>33%) with 51% above target. B) There is no specific pattern, with wide variability all day. Review of individual days shows the following issues. C) Fasting hypoglycemia occurs from too much correction dose at bedtime; miscalculated carbohydrate intake at breakfast (thought it was 15 g but was actually 35 g) resulted in noontime hyperglycemia that was not treated; correction dose insulin was not provided until 3:00 p.m., and appropriate prandial insulin was provided at 4:00 p.m. D) The patient missed 18 units of glargine the night before; mealtime and correction doses did not lower glucose normally without basal insulin. E) Patient received a good basal insulin dose (18 units glargine), keeping glucose flat overnight, but her 8:00 a.m. coffee resulted in a spike; she needs to give prandial insulin even for black coffee.

FIGURE 2.

AGP report notes for a 24-year-old woman who has had type 1 diabetes for 2 years and is beginning FCGM. Her A1C is 7.8%. A) Statistical summary shows high average glucose and high CV (>33%) with 51% above target. B) There is no specific pattern, with wide variability all day. Review of individual days shows the following issues. C) Fasting hypoglycemia occurs from too much correction dose at bedtime; miscalculated carbohydrate intake at breakfast (thought it was 15 g but was actually 35 g) resulted in noontime hyperglycemia that was not treated; correction dose insulin was not provided until 3:00 p.m., and appropriate prandial insulin was provided at 4:00 p.m. D) The patient missed 18 units of glargine the night before; mealtime and correction doses did not lower glucose normally without basal insulin. E) Patient received a good basal insulin dose (18 units glargine), keeping glucose flat overnight, but her 8:00 a.m. coffee resulted in a spike; she needs to give prandial insulin even for black coffee.

Close modal

AGP With Personal and Professional FCGM

The AGP is a robust tool for educating patients about the effects of food choices, exercise, and medications on glucose levels (90). With minimal training, clinicians can look at the AGP to visualize and prioritize clinical problems and, through an ongoing process of shared decision-making with patients, introduce interventions to increase time in range without increasing hypoglycemia (40,91).

A basic report review should include time in range, as well as patterns of hypoglycemia, hyperglycemia, and prandial glucose excursions (91). Results should be shared face-to-face with patients, if possible, using the report (Figure 2) as a decision aid to draw connections between glucose data, medication timing and dosing, eating, and other factors affecting glucose change (41,92).Information such as medication regimen, meals, exercise, and snacking should be noted directly under the curve on the printed AGP sheet. Once the sheet is marked up, asking the patient to briefly describe what he or she sees as possible reasons for glycemic excursions often generates useful insights.

Actions should be prioritized according to patterns of hypoglycemia, hyperglycemia, and glucose variability. For example, if the 10% lower line is touching the 70 mg/dL target line—a signal that, at that time of day, 10% of all glucose levels are <70 mg/dL—steps should be taken to reduce hypoglycemia; if the light blue area is very wide, conveying high glycemic variability, factors that may contribute to glucose fluctuations, such as the timing or amount of food intake, timing or dosing of medications, or patterns of exercise, should be considered for each time period. Questions should include:

  • Are pre-meal glucose levels at target?

  • Do glucose levels rise or fall dramatically after eating?

  • Do upward or downward fluctuations occur overnight?

  • Is there an explanation for glucose variability?

  • What typically happens with exercise?

  • Are patterns different on week-ends?

  • Are there circumstances such as stress or unusual travel that require special attention?

  • Is A1C consistent with daily glucose management?

At the end of the consultation, the key lessons of the AGP analysis should be summarized so that the patient comes away with one or two achievable recommendations (91). In all cases, the first priority should be treating hypoglycemia, indicated by the blue curves touching the 70 mg/dL line or lower. Timely follow-up is important to determine next steps, with a shorter interim for therapy adjustment than for lifestyle change recommendations.

Educators have an important role to play in encouraging wider acceptance of CGM, used either intermittently or for everyday diabetes management. Success will depend in part on dispelling long-held preconceptions based on traditional devices (44). Using both FCGM and the AGP, educators can more easily integrate retrospective continuous data analysis into busy practice settings and support patients so that they gain confidence in their ability to translate on-demand continuous data into appropriate action. FCGM, as an affordable option embraced by the international patient community with minimal training, may point the way to feasibility of CGM for more patients in the United States (40). Figure 3 provides a case example of an individual with type 2 diabetes who increased time in range and reduced A1C based on FCGM guidance alone.

FIGURE 3.

A) AGP profile and B) follow-up summary report of a 60-year-old man with type 2 diabetes. He was diagnosed with diabetes 3 years ago. His physician prescribed metformin and canagliflozin. With an A1C usually between 6.5 and 7.5%, and 42% of values within the hyperglycemic range as of August 2017, he began using personal FCGM the following December. His medication therapy remained the same. When he returned to the clinic in April 2018, 97% of his glucose values were within target range, and his estimated A1C was 5.8%. He attributed this improvement to controlling his rice intake based on feedback from FCGM.

FIGURE 3.

A) AGP profile and B) follow-up summary report of a 60-year-old man with type 2 diabetes. He was diagnosed with diabetes 3 years ago. His physician prescribed metformin and canagliflozin. With an A1C usually between 6.5 and 7.5%, and 42% of values within the hyperglycemic range as of August 2017, he began using personal FCGM the following December. His medication therapy remained the same. When he returned to the clinic in April 2018, 97% of his glucose values were within target range, and his estimated A1C was 5.8%. He attributed this improvement to controlling his rice intake based on feedback from FCGM.

Close modal

Clinical trial and empirical evidence attest that continuous glucose data analysis, used as an adjunct to A1C, provides more instructive and actionable information than SMBG. Although CGM is recognized as a robust tool for refining diabetes management, cost and reimbursement issues, scant resources to learn or implement new technology, and human factors of usability have hampered its wider use (15,28).

FCGM offers an easy, intuitive, and flexible option for appropriate patients with type 1 or type 2 diabetes (64,68). Candidates include people with newly diagnosed diabetes using SMBG with mixed success, as well as those on intensive insulin regimens who do not have impaired awareness of hypoglycemia.

Real-world data associate higher rates of FCGM scanning with improved glycemic markers, including increased time in range and reduced time in hyper- and hypoglycemia, and further evaluation is expected to build momentum toward incorporating continuous data strategies into daily self-care (48). Alternatively, professional FCGM offers a low-cost and nearly burden-free opportunity to understand patterns of high and low glucose that can be addressed in a timely fashion.

I.B.H. has received research support from Medtronic Diabetes and consulting income from Abbott Diabetes Care, Becton Dickinson, Bigfoot, and Roche. E.N. has received honoraria for participating on the speakers bureaus of Astra Zeneca and Janssen Pharmaceuticals and consulting income from Senseonics. No other potential conflicts of interest relevant to this article were reported.

Abbott Diabetes Care provided funding to The Diabetes Education Group in support of the writing of this article.

I.B.H., E.N., and C.A.V. developed the manuscript and reviewed its content. I.B.H. is the guarantor of this work and, as such, takes full responsibility for its integrity and accuracy.

1.
Rubinow
KB
,
Hirsch
IB
.
Reexamining metrics for glucose control
.
JAMA
2011
;
305
:
1132
1133
2.
American Diabetes Association
.
6. Glycemic targets: Standards of Medical Care in Diabetes—2018
.
Diabetes Care
2018
;
41
(
Suppl. 1
):
S55
S64
3.
Polonsky
WH
,
Fisher
L
,
Schikman
CH
, et al
.
Structured self-monitoring of blood glucose significantly reduces A1C levels in poorly controlled, noninsulin-treated type 2 diabetes: results from the Structure Testing Program study
.
Diabetes Care
2011
;
34
:
262
267
4.
Blevins
T
.
Value and utility of self-monitoring of blood glucose in non-insulin-treated patients with type 2 diabetes mellitus
.
Postgrad Med
2013
;
125
:
191
204
5.
Polonsky
WH
,
Fisher
L
,
Hessler
D
,
Edelman
SV
.
What is so tough about self-monitoring of blood glucose? Perceived obstacles among patients with type 2 diabetes
.
Diabetic Med
2014
: 
31
:
40
46
6.
Austin
MM
.
The two skill sets of self-monitoring of blood glucose education: the operational and the interpretive
.
Diabetes Spectr
2013
;
26
:
83
90
7.
Centers for Disease Control and Prevention. National Diabetes Statistics Report
,
2017. Atlanta, Ga., Centers for Disease Control and Prevention
,
U.S. Dept of Health and Human Services
,
2017
8.
Ali
MK
,
Bullard
KM
,
Saaddine
JB
,
Cowie
CC
,
Iperatore
G
,
Gregg
EW
.
Achievement of goals in US diabetes care
.
N Engl J Med
2013
;
368
:
1613
1624
9.
Carls
G
,
Huynh
J
,
Tuttle
E
,
Yee
J
,
Edelman
SV
.
Achievement of glycated hemoglobin foals in the US remains unchanged through 2014
.
Diabetes Ther
2017
;
8
:
863
873
10.
American Diabetes Association
.
1. Improving care and promoting health in populations: Standards of Medical Care in Diabetes—2018
.
Diabetes Care
2018
;
41
(
Suppl. 1
):
S7
S12
11.
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group
.
Effectiveness of continuous glucose monitoring in a clinical care environment: evidence from the Juvenile Diabetes Research Foundation Continuous Glucose Monitoring (JDRF-CGM) trial
.
Diabetes Care
2010
;
33
:
17
22
12.
Hermanides
J
,
Nørgaard
K
,
Bruttomesso
D
, et al
.
Sensor-augmented pump therapy lowers HbA1c in suboptimally controlled type 1 diabetes; a randomized controlled trial
.
Diabet Med
2011
;
28
:
1158
1167
13.
Battelino
T
,
Conget
I
,
Olsen
B
, et al.;
SWITCH Study Group
.
The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial
.
Diabetologia
2012
;
55
:
3155
3162
14.
New
JP
,
Ajjan
R
,
Pfeiffer
AF
,
Freckmann
G
.
Continuous glucose monitoring in people with diabetes: the randomized controlled Glucose Level Awareness in Diabetes Study (GLADIS)
.
Diabet Med
2015
;
32
:
609
617
15.
Wong
JC
,
Foster
NC
,
Maahs
DM
, et al
.;
T1D Exchange Clinic Network. Real-time continuous glucose monitoring among participants in the T1D Exchange clinic registry
.
Diabetes Care
2014
;
37
:
2702
270
16.
Riveline
JP
,
Schaepelynck
P
,
Chaillous
L
, et al
.;
EVADIAC Sensor Study Group. Assessment of patient-led or physician-driven continuous glucose monitoring in patients with poorly controlled type 1 diabetes using basal-bolus insulin regimens: a 1-year multicenter study
.
Diabetes Care
2012
;
35
:
965
971
17.
Rachmiel
M
,
Landau
Z
,
Boaz
M
, et al
.
The use of continuous glucose monitoring systems in a pediatric population with type 1 diabetes mellitus in real-life settings: the AWeSoMe Study Group experience
.
Acta Diabetol
2015
;
52
:
323
329
18.
Bergenstal
RM
,
Klonoff
DC
,
Garg
SK
, et al
.;
ASPIRE In-Home Study Group. Threshold-based insulin-pump interruption for reduction of hypoglycemia
.
N Engl J Med
2013
;
369
:
224
232
19.
Weinstock
RS
,
Xing
D
,
Maahs
DM
, et al.;
T1D Exchange Clinic Network
.
Severe hypoglycemia and diabetic ketoacidosis in adults with type 1 diabetes: results from the T1D Exchange clinic registry
.
J Clin Endocrinol Metab
2013
;
98
:
3411
3419
20.
Pickup
JC
,
Freeman
SC
,
Sutton
AJ
.
Glycaemic control in type 1 diabetes during real time continuous glucose monitoring compared with self monitoring of blood glucose: meta-analysis of randomised controlled trials using individual patient data
.
BMJ
2011
;
343
:
d3805
21.
Garg
SK
,
Voelmle
MK
,
Beatson
CR
, et al
.
Use of continuous glucose monitoring in subjects with type 1 diabetes on multiple daily injections versus continuous subcutaneous insulin infusion therapy: a prospective 6-month study
.
Diabetes Care
2011
;
34
:
574
579
22.
Beck
RW
,
Riddlesworth
T
,
Ruedy
K
, et al.;
DIAMOND Study Group
.
Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial
.
JAMA
2017
;
317
:
371
378
23.
Fonseca
VA
,
Grunbrger
G
,
Anhalt
H
, et al
.
Continuous glucose monitoring: a consensus conference of the American Association of Clinical Endocrinologists and American College of Endocrinology
.
Endocr Pract
2016
;
22
:
1008
1021
24.
T1D Exchange Clinic Network
.
Real-time continuous glucose monitoring among participants in the T1D Exchange clinic registry
.
Diabetes Care
2014
;
37
:
2702
2709
25.
Miller
KM
,
Foster
NC
,
Beck
RW
, et al.;
T1D Exchange Clinical Network
.
Current state of type 1 diabetes treatment in the U.S: updated data from the T1D Exchange clinic registry
.
Diabetes Care
2015
;
38
:
971
978
26.
Anhalt
H
.
Limitations of continuous glucose monitor usage
.
Diabetes Technol Ther
2016
;
18
:
115
117
27.
T1D Exchange
.
Why do some people with T1D stop using a pump and CGM?
[Internet] Available from t1dexchange.org/pages/why-do-some-people-with-t1d-stop-using-a-pump-and-cgm. Accessed 31 February 2018
28.
Petrie
JR
,
Peters
AL
,
Bergenstal
RM
,
Hall
RW
,
Fleming
AG
,
Heinemann
L
.
Improving the clinical value and utility of CGM systems: issues and recommendations: a joint statement of the European Association for the Study of Diabetes and the American Diabetes Association diabetes technology workshop
.
Diabetes Care
2017
;
40
:
1614
1621
29.
Tanenbaum
ML
,
Hanes
SJ
,
Miller
KM
,
Naranjo
D
,
Bensen
R
,
Hood
KK
.
Diabetes device use in adults with type 1 diabetes: barriers to uptake and potential intervention targets
.
Diabetes Care
2017
;
40
:
181
187
30.
Danne
T
,
Nimri
R
,
Battelino
T
, et al
.
International consensus on use of continuous glucose monitoring
.
Diabetes Care
2017
;
40
:
1631
1640
31.
U.S. Food and Drug Administration
.
FDA approves first continuous glucose monitoring system for adults not requiring blood sample calibration
.
27 September 2017. Available from www.fda.gov/NewsEvents/Newsroom/PressAnnouncements/ucm577890.htm. Accessed 14 March 2018
32.
Blum
A
.
Freestyle Libre glucose monitoring system
.
Clin Diabetes
2018
;
36
:
203
204
33.
U.S. Food and Drug Administration
.
Approval order: Freestyle Libre Pro Flash Glucose Monitoring System. PI50021. Silver Spring, Md., U.S. Department of Health and Human Services
,
2016
.
34.
Leelarathna
L
,
Wilmot
EG
.
Flash forward: a review of flash glucose monitoring
.
Diabet Med
2018
;
35
:
472
482
35.
Close
K
,
Brown
A
.
Abbott’s FreeStyle Libre transforming glucose monitoring through utter simplicity, fingersticks aside! Diatribe January 9, 2015. Available from diatribe.org/abbott-freestyle-libre-transforming-glucose-monitoring-through-utter-simplicity-fingersticks. Accessed 3 March 2018
36.
Schnell
O
,
Alawi
H
,
Battelino
T
, et al
.
Self-monitoring of blood glucose in type 2 diabetes: recent studies
.
J Diabetes Sci Technol
2013
;
7
:
478
488
37.
Vigersky
RA
,
Fonda
SJ
,
Chellappa
M
,
Walker
S
,
Ehrhardt
NM
.
Real-time continuous glucose monitoring in patients with type 2 diabetes
.
Diabetes Care
2012
;
35
:
32
38
38.
Vigersky
R
,
Shrivastav
M
.
Role of continuous glucose monitoring for type 2 in diabetes management and research
.
J Diabetes Complications
2017
;
31
:
280
287
39.
Heinemann
L
,
Freckmann
G
.
CGM versus FGM; or, continuous glucose monitoring is not flash glucose monitoring
.
J Diabetes Sci Technol
2015
;
9
:
947
950
40.
Distiller
La
,
Cranston
I
,
Mazze
R
.
First clinical experience with retrospective flash glucose monitoring (FGM) analysis in South Africa: characterizing glycemic control with ambulatory glucose profile
.
J Diabetes Sci Technol
2016
;
10
:
1294
1302
41.
Hirsch
IB
,
Verderese
CA
.
Professional flash continuous glucose monitoring with ambulatory glucose profile reporting to supplement A1C: rationale and practical implementation
.
Endocr Pract
2017
;
23
:
1333
1344
42.
Abbott Diabetes Care
.
FreeStyle Pro Operator’s Manual. Alameda, Calif., Abbott Diabetes Care
,
2017
.
Available from provider.myfreestyle.com/freestyle-libre-pro-resources.html. Accessed 6 December 2018
43.
American Association of Diabetes Educators
.
White paper: continuous glucose monitoring summit
.
Chicago, Ill
.,
American Association of Diabetes Educators
,
2015
.
44.
Rodbard
D
.
Continuous glucose monitoring: a review of successes, challenges, and opportunities
.
Diabetes Technol Ther
2016
;
18
(
Suppl. 2
):
S3
S13
45.
Blevins
TC
,
Bode
BW
,
Garg
SK
, et al
.
Statement by the American Association of Clinical Endocrinologists Consensus Panel on Continuous Glucose Monitoring
.
Endocr Pract
2010
;
16
:
730
745
46.
Adolfsson
P
,
Parkin
CG
,
Thomas
A
,
Krinelke
LG
.
Selecting the appropriate monitoring system: a practical approach
.
Eur Endocrinol
2018
;
14
:
24
29
47.
Hirsch
IB
,
Battelino
T
,
Peters
AL
,
Chamberlain
JJ
,
Aleppo
G
,
Bergenstal
RM
.
Role of Continuous Glucose Monitoring in Diabetes Treatment
.
Arlington, Va
.,
American Diabetes Association
,
2018
48.
Dunn
TC
,
Xu
Y
,
Hayter
G
,
Ajjan
RA
.
Real-world flash glucose monitoring patterns associated between self-monitoring frequency and glycaemic measures: a European analysis of over 60 million glucose tests
.
Diabetes Res Clin Pract
2018
;
137
:
37
46
49.
Fokkert
MJ
,
van Dijk
PR
,
Edens
MA
, et al
.
Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and type 2 diabetes mellitus
.
BMJ Open Diabetes Res Care
2017
;
5
:
e000320
50.
Centers for Medicare & Medicaid Services
.
Medicare coverage of diabetes supplies and services
.
51.
Maahs
DM
,
DeSalvo
D
,
Pyle
L
, et al
.
Effect of acetaminophen on CGM glucose in an outpatient setting
.
Diabetes Care
2015
;
38
:
e158
e159
52.
Polonsky
WH
,
Hessler
D
.
Perceived accuracy in continuous glucose monitoring: understanding the impact on patients
.
J Diabetes Sci Technol
2015
;
9
:
330
341
53.
Edelman
S
,
Pettus
J
,
Price
D
.
How current continuous glucose monitoring (CGM) users translate CGM data into diabetes management: results of a survey
.
Diabetologia
2014
;
57
(
Suppl 1
):
1020
54.
Clarke
W
,
Kovatchev
B
.
Statistical tools to analyze continuous glucose monitor data
.
Diabetes Technol Ther
2009
;
11
(
Suppl. 1
):
S45
S54
55.
Bailey
T
,
Bode
BW
,
Christiansen
MP
, et al
.
The performance and usability of a factory-calibrated flash glucose monitoring system
.
Diabetes Technol Ther
2015
;
17
:
787
794
56.
U.S. Food and Drug Administration
.
Summary of safety and effectiveness data (SSED). Dexcom G4 PLATINUM continuous glucose monitoring system
.
Available from www.accessdata.fda.gov/cdrh_docs/pdf12/P120005b.pdf. Accessed 23 April 2018
57.
U.S. Food and Drug Administration
.
Summary of safety and effectiveness data (SSED)
.
Medtronic MiniMed 530G (Enlite sensor). Available from www.accessdata.fda.gov/cdrh_docs/pdf15/P150019b.pdf. Accessed 23 April 2018
58.
U.S. Food and Drug Administration
.
Summary of safety and effectiveness data (SSED). FreeStyle Libre Flash glucose monitoring system
.
Available from www.accessdata.fda.gov/cdrh_docs/pdf16/P160030B.pdf. Accessed 23 April 2018
59.
Aberer
F
,
Hajnsek
M
,
Rumpler
M
, et al
.
Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes
.
Diabetes Obes Metab
2017
;
19
:
1051
1055
60.
Bonora
B
,
Maran
A
,
Ciciliot
S
,
Avogaro
A
,
Fadini
GP
.
Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetes
.
J Endocrinol Invest
2016
;
39
:
1391
1399
61.
Boscari
F
,
Galasso
S
,
Facchinetti
A
, et al
.
FreeStyle Libre and Dexcom G4 Platinum sensors: accuracy comparisons during two weeks of home use and use during experimentally induced glucose excursions
.
Nutr Metab Cardiovasc Dis
2018
;
28
:
180
186
62.
Cobelli
C
,
Schiavon
M
,
Dalla Man
C
,
Basu
A
,
Basu
R
.
Interstitial fluid glucose is not just a shifted-in-time but a distorted mirror of blood glucose: insight from an in silico study
.
Diabetes Technol Ther
2016
;
18
:
505
511
63.
Boscari
F
,
Galasso
S
,
Acciaroli
G
, et al
.
Head-to-head comparison of the accuracy of Abbott FreeStyle Libre and Dexcom G5 mobile
.
Nutr Metab Cardiovasc Dis
2018
;
28
:
425
427
64.
Bolinder
J
,
Antuna
R
,
Geelhoed-Duijvestijn
P
,
Kroger
J
,
Weitgasser
R
.
Novel glucose-sensing technology and hypoglycemia in type 1 diabetes: a multicentre, non-masked, randomized controlled trial
.
Lancet
2016
;
388
:
2254
2263
65.
Olafsdottier
AF
,
Attvall
S
,
Sandgren
U
, et al
.
A clinical trial of the accuracy and treatment experience of the flash glucose monitor FreeStyle Libre in adults with type 1 diabetes
.
Diabetes Technol Ther
2017
;
19
:
164
172
66.
Dover
AR
,
Stimson
RH
,
Zammit
NN
,
Gib
FW
.
Flash glucose monitoring improves outcomes in a type 1 diabetes clinic
.
J Diabetes Sci Technol
2017
;
11
:
442
443
67.
Ish-Shalom
M
,
Wainstein
J
,
Raz
I
,
Mosenzon
O
.
Improvement in glucose control in difficult-to-control patients with diabetes using a novel flash glucose monitoring device
.
J Diabetes Sci Technol
2016
;
10
:
1412
1413
68.
Haak
T
,
Hanaire
H
,
Ajjan
R
,
Hermanns
N
,
Riveline
JP
,
Rayman
G
.
Flash glucose-sensing technology as a replacement for blood glucose monitoring for the management of insulin-treated type 2 diabetes: a multicenter, open-label randomized controlled trial
.
Diabetes Ther
2017
;
8
:
55
73
69.
Berwick
DM
,
Nolan
TW
,
Whittington
J
.
The triple aim: care, health, and cost
.
Health Aff (Millwood)
2008
;
27
:
759
769
70.
Nathan
DM
,
Kuenen
J
,
Borg
R
,
Zhen
H
,
Schoenfeld
D
,
Heine
RJ
;
A1c-Derived Average Glucose Study Group. Translating the A1C assay into estimated average glucose values
.
Diabetes Care
2008
;
31
:
1473
1478
71.
Wright
LA
,
Hirsch
HB
.
Metrics beyond hemoglobin A1C in diabetes management: time in range, hypoglycemia, and other parameters
.
Diabetes Technol Ther
2017
;
19
(
Suppl. 2
):
S16
S26
72.
Beck
RW
,
Connor
CG
,
Mullen
D
,
Wesley
DM
,
Bergenstal
RM
.
The fallacy of average: how using A1C alone to assess glycemic control can be misleading
.
Diabetes Care
2017
;
40
:
994
999
73.
Bergenstal
RM
.
Glycemic variability and diabetes complications: does it matter? Simply put, there are better glycemic markers!
Diabetes Care
2015
;
38
:
1615
1621
74.
Hirsch
IB
,
Brownlee
M
.
Beyond hemoglobin A1c: need for additional markers of risk for diabetic microvascular complications
.
JAMA
2010
;
303
:
2291
2292
75.
Cryer
PE
.
Severe hypoglycemia predicts mortality in diabetes
.
Diabetes Care
2012
;
35
:
1814
1816
76.
Vigersky
RA
.
Escaping the hemoglobin A1c-centric world in evaluating diabetes mellitus interventions
.
J Diabetes Sci Technol
2015
;
9
:
1148
1151
77.
Monnier
L
,
Mas
E
,
Ginet
C
, et al
.
Activation of oxidative stress by acute glucose fluctuations compared with sustained chronic hyperglycemia in patients with type 2 diabetes
.
JAMA
2006
;
295
:
1681
1687
78.
Esposito
K
,
Ciotola
M
,
Carleo
D
, et al
.
Post-meal glucose peaks at home associate with carotid intima-media thickness in type 2 diabetes
.
J Clin Endocrinol Metab
2008
;
93
:
1345
1350
79.
Monnier
L
,
Colette
C
,
Leiter
L
, et al
.
The effect of glucose variability on the risk of microvascular complications in type 1 diabetes
.
Diabetes Care
2007
;
30
:
185
188
80.
Qu
Y
,
Jacober
SJ
,
Zhang
Q
, et al
.
Rate of hypoglycemia in insulin-treated patients with type 2 diabetes can be predicted from glycemic variability data
.
Diabetes Technol Ther
2012
;
14
:
1008
1012
81.
Monnier
L
,
Wojtusciszyn
A
,
Colette
C
, et al
.
The contribution of glucose variability to asymptomatic hypoglycemia in persons with type 2 diabetes
.
Diabetes Technol Ther
2011
;
13
:
813
818
82.
Ceriello
A
,
Esposito
K
,
Piconi
L
, et al
.
Oscillating glucose is more deleterious to endothelial function and oxidative stress than mean glucose in normal and type 2 diabetic patients
.
Diabetes
2008
;
57
:
1349
1354
83.
Testa
MA
,
Gill
J
,
Su
M
, et al
.
Comparative effectiveness of basal-bolus versus premix analog insulin on glycemic variability and patient-centered outcomes during insulin intensification in type 1 and type 2 diabetes: a randomized, controlled, crossover trial
.
J Clin Endocrinol Metab
2012
;
97
:
3504
3514
84.
Cryer
PE
.
Glycemic goals in diabetes: trade-off between glycemic control and iatrogenic hypoglycemia
.
Diabetes
2014
;
63
:
2188
2195
85.
Abbott Diabetes Care
.
FreeStyle Libre Flash Glucose Monitoring System User’s Manual. Alameda, Calif., Abbott Diabetes Care
,
2018
.
Available from www.accessdata.fda.gov/cdrh_docs/pdf16/P160030S017C.pdf. Accessed 6 December 2018
86.
Chamberlain
J
,
Dopita
D
,
Gilgen
E
.
Persistence of continuous monitoring use in a community setting 1 year after purchase
.
Clin Diabetes
2013
;
31
:
106
109
87.
Kudva
YC
,
Ahmann
AJ
,
Bergenstal
RM
, et al
.
Approach to using trend arrows in the FreeStyle Libre Flash glucose monitoring systems in adults
.
J Endocr Soc
2018
;
2
:
1320
1337
88.
Bergenstal
RM
,
Ahmann
AJ
,
Bailey
T
, et al
.
Recommendations for standardizing glucose reporting and analysis to optimize clinical decision making in diabetes: the Ambulatory Glucose Profile (AGP)
.
Diabetes Technol Ther
2013
;
15
:
198
211
89.
Mazze
RS
,
Lucido
D
,
Langer
O
,
Hartmann
K
,
Rodbard
D
.
Ambulatory glucose profile: representation of verified self-monitored blood glucose data
.
Diabetes Care
1987
;
10
:
111
117
90.
Allen
NA
,
Fain
JA
,
Braun
B
, et al
.
Continuous glucose monitoring in non-insulin-using individuals with type 2 diabetes: acceptability, feasibility, and teaching opportunities
.
Diabetes Technol Ther
2009
;
11
:
151
158
91.
Carlson
AL
,
Mullen
DM
,
Bergenstal
RM
.
Clinical use of continuous monitoring in adults with type 2 diabetes
.
Diabetes Ther
2017
;
19
(
Suppl. 2
):
S4
S11
92.
Rodriguez-Gutierrez
R
,
Gionfriddo
MR
,
Ospina
NS
, et al
.
Shared decision making in endocrinology: present and future directions
.
Lancet Diabetes Endocrinol
2016
;
4
:
706
716
Readers may use this article as long as the work is properly cited, the use is educational and not for profit, and the work is not altered. See www.diabetesjournals.org/content/license for details.