Table 1

Difference between GMI (calculated from CGM-derived mean glucose) and laboratory-measured HbA1c

Absolute value of difference between GMI and laboratory HbA1c (%)Percentage of values (95% CI)
Guardian 3 sensor (n = 85)Navigator 2 (n = 114)Dexcom sensors**
0 to <0.1 19 (11–29) 20 (13–29) 19 (16–22) 
≥0.1 81 (71–89) 80 (72–87) 81 (78–84) 
≥0.2 66 (55–76) 68 (58–76) 67 (63–71) 
≥0.3 54 (43–66) 56 (46–65) 51 (47–55) 
≥0.4 42 (32–54) 46 (36–56) 39 (34–43) 
≥0.5 32 (22–43) 36 (27–46) 28 (24–32) 
≥0.6 24 (15–34) 28 (20–37) 19 (15–22) 
≥0.7 13 (7–22) 21 (14–30) 12 (9–15) 
≥0.8 11 (5–19) 12 (6–19) 8 (5–10) 
≥0.9 5 (1–12) 8 (4–15) 4 (3–6) 
≥1 3 (1–10) 5 (2–10) 3 (2–4) 
Absolute value of difference between GMI and laboratory HbA1c (%)Percentage of values (95% CI)
Guardian 3 sensor (n = 85)Navigator 2 (n = 114)Dexcom sensors**
0 to <0.1 19 (11–29) 20 (13–29) 19 (16–22) 
≥0.1 81 (71–89) 80 (72–87) 81 (78–84) 
≥0.2 66 (55–76) 68 (58–76) 67 (63–71) 
≥0.3 54 (43–66) 56 (46–65) 51 (47–55) 
≥0.4 42 (32–54) 46 (36–56) 39 (34–43) 
≥0.5 32 (22–43) 36 (27–46) 28 (24–32) 
≥0.6 24 (15–34) 28 (20–37) 19 (15–22) 
≥0.7 13 (7–22) 21 (14–30) 12 (9–15) 
≥0.8 11 (5–19) 12 (6–19) 8 (5–10) 
≥0.9 5 (1–12) 8 (4–15) 4 (3–6) 
≥1 3 (1–10) 5 (2–10) 3 (2–4) 

**Dexcom data from Bergenstal et al. (1).