Table 7.4

Continuous glucose monitoring devices interfering substances

MedicationSystems affectedEffect
Acetaminophen   
 >4 g/day Dexcom G6 Higher sensor readings than actual glucose 
 Any dose Medtronic Guardian Higher sensor readings than actual glucose 
Alcohol Medtronic Guardian Sensor readings may be higher than actual glucose 
Ascorbic acid (vitamin C), >500 mg/day FreeStyle Libre Higher sensor readings than actual glucose 
Hydroxyurea Dexcom G6, Medtronic Guardian Higher sensor readings than actual glucose 
Mannitol Senseonics Eversense Sensor bias within therapeutic concentration ranges 
Tetracycline Senseonics Eversense Sensor bias within therapeutic concentration ranges 
MedicationSystems affectedEffect
Acetaminophen   
 >4 g/day Dexcom G6 Higher sensor readings than actual glucose 
 Any dose Medtronic Guardian Higher sensor readings than actual glucose 
Alcohol Medtronic Guardian Sensor readings may be higher than actual glucose 
Ascorbic acid (vitamin C), >500 mg/day FreeStyle Libre Higher sensor readings than actual glucose 
Hydroxyurea Dexcom G6, Medtronic Guardian Higher sensor readings than actual glucose 
Mannitol Senseonics Eversense Sensor bias within therapeutic concentration ranges 
Tetracycline Senseonics Eversense Sensor bias within therapeutic concentration ranges 